Diagnostic reference level curves for paediatric fluoroscopic imaging in the Netherlands
Goswin Croes1,2, Ingrid Nijholt3, Martijn Boomsma3
1Department of Medical Physics, Isala Hospital, Zwolle, The Netherlands. g.o.croes@isala.nl.
European Radiology
|September 6, 2025
Summary
Establishing updated paediatric fluoroscopic diagnostic reference levels (DRLs) is crucial. This study found current Dutch DRLs for micturating cystourethrography may be too high, suggesting a dose-age curve approach for better paediatric radiation dose optimisation.
Area of Science:
- Radiology
- Medical Physics
- Paediatric Imaging
Background:
- Establishing paediatric diagnostic reference levels (DRLs) is challenging due to limited data availability.
- Current European Guidelines on DRLs for Paediatric Imaging (PiDRL) necessitate updating existing DRLs.
- Paediatric fluoroscopic dose levels require assessment in clinical practice.
Purpose of the Study:
- To assess paediatric fluoroscopic dose levels in Dutch clinical practice.
- To update current Dutch diagnostic reference levels (DRLs) for paediatric fluoroscopy.
- To evaluate the utility of a dose-age curve method for establishing DRLs with sparse data.
Main Methods:
- Retrospective collection of Air Kerma-area Product (KAP) values from paediatric patients (0-18 years) undergoing fluoroscopic procedures (MCU, Upper GI, Lower GI) between 2017 and 2021.
- Application of PiDRL recommendations for sparse data: fitting the 75th percentiles of median KAP values per age group using an exponential dose-age curve.
- Comparison of proposed DRL values with existing Dutch and European DRLs.
Main Results:
- A total of 971 paediatric fluoroscopic examinations were analysed.
- The proposed DRL curve for micturating cystourethrography (MCU) was lower than current Dutch DRLs.
- The proposed DRL curve for upper gastrointestinal (Upper GI) procedures was lower than UK DRLs; no DRL curve could be established for lower gastrointestinal (Lower GI) due to insufficient data.
Conclusions:
- Paediatric fluoroscopic dose levels in the study are significantly lower than current Dutch DRLs, particularly for MCU.
- A dose-age curve method is proposed for updating Dutch paediatric DRLs, enabling DRL establishment even with sparse data.
- The proposed method facilitates the optimisation of radiation doses in paediatric patients.