Long-Term Performance of Passive Volatile Organic Compounds (VOCs) Samplers for Indoor Air
John H Zimmerman1, Brian Schumacher2, Christopher C Lutes3
1Watershed and Ecosystem Characterization Division, Center for Environmental Measurement & Modeling (CEMM), Office of Research and Development (ORD), U.S. Environmental Protection Agency (EPA), 109 TW Alexander Dr., Research Triangle Park, NC 27711, USA.
Passive samplers accurately measure indoor volatile organic compounds (VOCs) over extended periods. Different samplers and compounds show varying reliability for long-term air monitoring and vapor intrusion assessments.
Area of Science:
- Environmental Chemistry
- Air Quality Monitoring
- Analytical Chemistry
Background:
- Reliability of passive samplers for indoor volatile organic compounds (VOCs) measurement is crucial.
- Sampler performance depends on constant uptake rates under varying environmental conditions and exposure durations.
- Vapor intrusion (VI) assessments require accurate long-term air monitoring data.
Purpose of the Study:
- To evaluate the performance of passive samplers for measuring VOCs over extended exposure durations (1 week to 1 year).
- To assess the reliability of different passive sampler types, including Radiello®, SKC 575, and Waterloo Membrane Sampler™ (WMS™).
- To determine the suitability of passive samplers for long-term indoor air quality monitoring and vapor intrusion studies.
Main Methods:
- Phase 1: Evaluated charcoal-based, solvent-extracted passive samplers (Radiello® 130, white diffusive bodies) for 1 week to 1 year in a test house with known VI.
- Phase 2: Tested Radiello® 130 (yellow diffusive bodies), SKC 575 (with secondary diffusive cover), and WMS™ for 6 months and 1 year.
- Measured VOC concentrations and calculated %Bias against acceptance criteria (±30%).
Main Results:
- Chloroform %Bias exceeded ±30% after 4 weeks.
- Benzene, hexane, and trichloroethylene (TCE) were within ±30% bias for up to 3 months.
- Toluene and tetrachloroethylene (PCE) showed uniform uptake over 1 year.
- SKC 575 and Radiello® 130 (yellow) were acceptable for 6 months; WMS™ favored petroleum hydrocarbons.
- All tested samplers showed acceptable performance for specific compounds and conditions after 1 year.
Conclusions:
- Passive samplers can provide acceptable indoor air quality data for VOCs over exposure periods exceeding 30 days, up to one year for certain compounds.
- Sampler choice and compound volatility influence long-term monitoring reliability.
- Results support the use of passive samplers for long-term vapor intrusion assessments under defined conditions.