Clinical validation of using a commercial synthetic-computed tomography solution for brain MRI-only radiotherapy
Lamyaa Aljaafari1,2,3, Richard Speight2,4, David L Buckley1
1Leeds Institute of Cardiovascular & Metabolic Medicine (LICAMM), University of Leeds, Woodhouse, Leeds LS2 9JT, United Kingdom.
Background And Purpose:
MRI-only radiotherapy treatment planning (RTP) relies on synthetic-CT (sCT) images for dose calculation. This study evaluates the clinical feasibility of using a commercial sCT solution in brain RTP, MRCAT Brain, focusing on dosimetric accuracy and patient setup verification.
Method And Materials:
For dosimetric evaluation, 93 patients with brain cancer who were treated with volumetric modulated arc therapy (VMAT) using a CT/MRI fusion workflow were included. sCT images were generated using MRCAT Brain. The sCT images were rigidly co-registered to the CT images. The clinical plan produced on the CT was recalculated on the sCT. Dosimetric accuracy was assessed by comparing dose differences in dose volume histogram (DVH) statistics for the planning target volume (PTV) and organs at risk (OARs).For patient setup verification, 70 patients were included, and total of 572 cone beam CT (CBCT) registrations were performed with sCT and CT as reference images. The sCT matching accuracy was validated by comparing the translational and rotational differences between sCT-CBCT and CT-CBCT registrations.
Results:
The PTV mean dose difference between CT and sCT were 0.3 %, 0.4 %, and 0.2 % for D50%, D2%, and D98%, respectively. The OAR mean dose differences were less than 0.3 % for all OARs. 4 of 93 patients (4.3 %) showed gross dosimetric errors of greater than ± 2 %. 3/4 were caused by sCT error. For positioning verification, all results were between ± 1 mm and ± 1°.
Conclusion:
This study demonstrates the clinical feasibility of the MRCAT solution for brain MRI-only RTP, with dosimetric differences being clinically acceptable, along with submillimetre and sub-degree accuracy in patient setup verification.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...