Related Experiment Video
Updated: Jan 17, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Transfer of the SIR key comparison reference values from one ionization chamber to another
Carine Michotte1, Andrew K Pearce2, Maurice G Cox2
1Bureau International des Poids et Mesures, 92 310 Sèvres, France.
None:
In the framework of the CIPM MRA, the continuous BIPM.RI(II)-K1 key comparison named SIR (Système International de Référence) is based on an ionization chamber (IC1) in which more than 70 radionuclides were measured producing about 800 comparison results (equivalent activities Ae in kBq) since 1976. A second chamber (IC2), built to the same nominal design specification, is available to serve as a back-up should the first chamber fail; however, there will be small variations in the chamber such as in wall thicknesses and the tilt of the well. Because of these differences, the key comparison reference values (KCRVs) from IC1 cannot be applied directly to IC2, and to date only 55 independent measurements of 30 radionuclides are available for IC2. A methodology to transfer all existing key comparison reference values (KCRVs) to the back-up chamber IC2 was developed with the aim of minimizing the impact on the uncertainty of the degrees of equivalence. The selected methodology is based on a mathematical model utilizing an updated version of the SIRIC code previously developed by the current authors. This code takes as input the measured IC2 data and IC1 KCRVs transformed by a scaling function S and provides as output the modelled IC2 equivalent activities AeIC2,mod and the quotient AeIC2,mod/AeIC1,mod of modelled equivalent activities for each chamber. The IC2 KCRVs are calculated using KCRVIC2=KCRVIC1AeIC2,mod/AeIC1,mod. The optimum form of the scaling function S was explored and potential IC2 KCRVs and degrees of equivalence were calculated and are presented.
More Related Videos
Related Concept Videos
Ion-Exchange Chromatography
Ionization Energy
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Mass Spectrometers
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview