CF-PEEK vs. Titanium Dental Implants: Stress Distribution and Fatigue Performance in Variable Bone Qualities
Nurdan Polat Sağsöz1, Fahri Murat2, Sema Nur Sevinç Gül1
1Faculty of Dentistry, Atatürk University, 25240 Erzurum, Turkey.
Biomimetics (Basel, Switzerland)
|September 26, 2025
Summary
Carbon fiber-reinforced polyetheretherketone (CF-PEEK) dental implants show reduced stress shielding compared to titanium, offering better load transfer. However, CF-PEEK may have lower fatigue life, especially in low-density bone, requiring further optimization.
Area of Science:
- Biomaterials Science
- Dental Implantology
- Finite Element Analysis
Background:
- Titanium is the gold standard for dental implants, but its high stiffness can lead to stress shielding.
- Carbon fiber-reinforced polyetheretherketone (CF-PEEK) presents a potential alternative with biomechanical properties closer to bone.
Purpose of the Study:
- To compare the biomechanical behavior of titanium and CF-PEEK dental implants.
- To evaluate their performance under various bone densities and loading conditions using finite element analysis (FEA).
Main Methods:
- A finite element model of a mandibular molar implant system was created using titanium or CF-PEEK components.
- Simulations included four bone density configurations and vertical/oblique static loads (100 N).
- Analysis focused on stress, strain, and fatigue life.
Main Results:
- Titanium implants showed higher von Mises stress (>400 MPa) under oblique loading.
- CF-PEEK implants exhibited lower stress but higher strain, with more uniform stress distribution in surrounding bone.
- CF-PEEK components demonstrated reduced fatigue life, particularly in low-density bone under oblique loads.
Conclusions:
- CF-PEEK implants facilitate more physiological load transfer and reduce stress shielding compared to titanium.
- The structural reliability of CF-PEEK requires careful consideration, especially in compromised bone conditions.
- Further material and design optimization are needed for CF-PEEK dental implants.