Characterizing geometrical distortions in MRI for radiotherapy: Evaluating MR-SIM protocols and dose accuracy in
Mojtaba Barzegar1,2, Aram Rostami1, Abbass Yousef Mkanna1
1Radiation Oncology Department, National Center for Cancer Care and Research, Doha, Qatar.
Objective:
Magnetic resonance imaging (MRI) is crucial for diagnostic imaging and radiotherapy (RT) planning due to its superior soft tissue contrast. However, geometric distortions can affect treatment accuracy. This study evaluates the geometric accuracy of MRI protocols using a 1.5T MR-Sim scanner for RT and their dosimetric impacts.
Materials And Methods:
Geometric distortion was assessed using the CIRS 604-GS MR Image Distortion Phantom with 2152 control points across various MRI sequences. A central cubic structure and surrounding regions were analyzed, totaling 27 structures. Rigid registration in Eclipse V18.1 aligned MRI images to CT images, with verified accuracy and visual inspection. An evaluation of the overlap and surface distance metrics was performed through MIM Mastro. Geometric distortion was quantified using 3D distortion analysis software, comparing marker coordinates to reference CT scans. Twenty-seven VMAT full arc plans were done in Eclipse on 27 paired structures, irradiated with 16 Gy in a single fraction at 6 MV energy using the Acuros XB algorithm RESULTS: All 3D sequences demonstrated mean distortions of 0.4-0.5 mm, with maximum distortions of up to 3.3 mm in the TOF (time of flight) Angiography sequence. In contrast, the T2 TSE 2D sequence showed larger distortions (mean 1.4 mm, max. 9.7 mm). The center structure showed stable dosimetric performance (Dmean: 1601.6 cGy for CT, 1600.3 cGy for MR). Peripheral regions showed higher variability, with MRnormalized D95% values ranging from 83.9% to 88.1%.
Conclusion:
Precise MR protocols are essential for accurate tumor delineation and RT planning. While all sequences showed acceptable accuracy, 3D sequences are superior for high-precision RT.
More Related Videos
10:48PET and MRI Guided Irradiation of a Glioblastoma Rat Model Using a Micro-irradiator
Published on: December 28, 2017
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging