Simulation of dB/dt-Over-Electric Field Cardiac Magnetostimulation Safety Ratios in 75 Body Models and 18 Gradient Systems
Related Experiment Video
Updated: May 7, 2026

Patient-specific Modeling of the Heart: Estimation of Ventricular Fiber Orientations
Published on: January 8, 2013
Valerie Klein1,2, Jonathan Edmonson3, Mathias Davids1
1A. A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, Massachusetts, USA.
Purpose:
The IEC 60601-2-33 standard provides consensus-based safety provisions for MRI equipment. Protection of patients against cardiac stimulation (CS) is based on limiting the maximum E-field induced by MRI gradient coils. In practice, this is achieved by imposing a conservative dB/dt threshold on any gradient waveform. The dB/dt-over-E-field conversion ratio currently used in IEC 60601-2-33 was derived in a homogeneous ellipsoid exposed to a uniform B-field and is 10 (T/s)*(V/m)-1. This limit is becoming increasingly restrictive in high performance clinical systems. We therefore evaluate dB/dt-over-E-field ratios in realistic body models and coils using state-of-the-art electromagnetic simulations.
Methods:
We performed two independent simulation studies in a total of 75 realistic body models and 13 commercial gradient systems and derived dB/dt-over-E-field ratios in the heart. We thresholded the E-field maps to mitigate the impact of staircasing artifacts in boundary voxels between the myocardium and the lungs.
Results:
Thresholding the E-field maps at the 99th percentile E-field value (E99) eliminates staircasing artifacts in both simulation studies. Study #1 predicts a larger range of dB/dt-over-E99 ratios (13-53 (T/s)*(V/m)-1) than study #2 (12-35 (T/s)*(V/m)-1). Despite differences in EM solvers, body models, coils, mesh resolution, and post-processing, both studies find similar worst-case ratios of dB/dt-over-E99 of 12-13 (T/s)*(V/m)-1.
Conclusion:
Our simulations of dB/dt-over-E-field ratios for cardiac safety in MRI cover a large range of realistic clinical scenarios. An increase of the allowable dB/dt beyond the current CS limit in IEC 60601-2-33 may be feasible.