Comparative evaluation of quantitative analysis using two phantom-based software programs for 18F-FDG-PET/CT
Kosuke Yamashita1, Kazuki Motegi2, Noriaki Miyaji3
1Department of Medical Imaging Technology, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan. kyamashita@kumamoto-u.ac.jp.
A new automated software, Arimaru, shows strong concordance with conventional methods for Fluorine-18-fluorodeoxyglucose positron emission tomography/computed tomography (18F-FDG-PET/CT) quantitative analysis, meeting Japanese Society of Nuclear Medicine guidelines.
Area of Science:
- Nuclear Medicine
- Medical Imaging
- Radiopharmacology
Background:
- Variability in 18F-FDG-PET/CT image quality and quantitative accuracy exists across institutions and devices.
- The Japanese Society of Nuclear Medicine (JSNM) established guidelines for image quality assurance.
- Automated software solutions are emerging to streamline quality assurance processes.
Purpose of the Study:
- To compare the performance of a new automated software, Arimaru, against conventional software for quantitative analysis of 18F-FDG-PET/CT images.
- To evaluate Arimaru's accuracy in calculating physical parameters relevant to image quality.
- To assess Arimaru's compliance with JSNM guidelines.
Main Methods:
- Utilized the NEMA IEC Body phantom with list-mode acquisition (1800s) and reconstruction at multiple time points (30-1800s) on two PET/CT systems (GE Healthcare Discovery MI and IQ).
- Calculated five physical parameters (Q H,10, N 10, Q H,10/N 10, CV BG, and SUV max) using both Arimaru and conventional software (PETquact IE).
- Assessed region of interest (ROI) positioning accuracy and correlation between methods.
Main Results:
- Arimaru accurately positioned ROIs and demonstrated strong correlations (r > 0.85, p < 0.05) with the conventional method for all parameters.
- Significant differences were observed in some physical parameter values between Arimaru and conventional software.
- The automated method met JSNM guidelines.
Conclusions:
- The automated Arimaru software shows strong concordance with conventional methods and adheres to JSNM guidelines for 18F-FDG-PET/CT quantitative analysis.
- Systematic differences in calculated values necessitate understanding software-specific characteristics.
- Adoption of automated tools like Arimaru can promote standardized PET imaging practices.