Leak during term neonatal CPAP stabilisation: does resuscitation device design affect delivered PEEP - a bench study
Viktoria Gruber1,2, Stephanie Morakeas3,4, Murray Hinder4,5
1Department of Paediatrics and Adolescent Medicine, Medical University of Graz Division of Neonatology, Graz, Austria.
BMJ Paediatrics Open
|October 16, 2025
Summary
Mask leak during newborn resuscitation impacts ventilation. The rPAP device maintained continuous positive airway pressure (CPAP) better than Neopuff during leaks, crucial for effective neonatal respiratory support.
Area of Science:
- Neonatal respiratory support
- Mechanical ventilation
- Cardiopulmonary resuscitation
Background:
- Mask leak is a frequent challenge during neonatal resuscitation, potentially compromising ventilation efficacy.
- The impact of mask leak on continuous positive airway pressure (CPAP) delivery during assisted spontaneous breathing in neonates is not well-understood.
Purpose of the Study:
- To compare the performance of two T-piece resuscitation devices, the high-resistance Neopuff and the low-resistance rPAP, in the presence of mask leak during CPAP delivery.
- To evaluate how differing device resistances affect CPAP delivery under various leak conditions.
Main Methods:
- A term dynamic lung model was utilized to simulate neonatal respiratory mechanics.
- Two T-piece devices (Neopuff and rPAP) were tested at varying set positive end-expiratory pressure (PEEP) levels (5, 7, and 9 cmH2O) with simulated continuous leaks.
- Key parameters including mean airway pressure, pressure fluctuations, tidal volume (VT), and leak flow were measured.
Main Results:
- Mask leak significantly reduced delivered PEEP, with greater reductions observed at higher PEEP settings and leak levels.
- The rPAP device maintained delivered PEEP closer to the set value compared to the Neopuff device across all leak conditions.
- Tidal volume delivery was reduced with the Neopuff device, particularly at lower leak levels, and was influenced by the delivered PEEP.
Conclusions:
- The performance of T-piece resuscitation devices in the presence of mask leak varies significantly.
- The rPAP device demonstrated superior ability to maintain target PEEP levels during leaks compared to the Neopuff.
- Device resistance influences the impact of leaks on delivered tidal volume and overall ventilatory support in neonates.
More Related Videos
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
508
Noninvasive positive-pressure ventilation (NIPPV), continuous positive airway pressure (CPAP), and bilevel positive airway pressure (BiPAP) are essential methods in respiratory care. These ventilation techniques offer unique benefits for patients with various respiratory conditions, providing adequate support without requiring intubation. Let's explore how each method is crucial in improving patient outcomes and enhancing respiratory therapy.
Noninvasive Positive-Pressure Ventilation...
Noninvasive Positive-Pressure Ventilation...
508
Cardiopulmonary Resuscitation II: ACLS Airway Management
514
Airway management is a key skill in emergency and critical care settings, as maintaining a clear airway is essential for adequate oxygenation and ventilation.Head Tilt-Chin Lift TechniqueThe head tilt-chin lift maneuver is an essential technique primarily used in patients without suspected cervical spine injuries. To perform this maneuver, one hand is placed on the patient’s forehead, and gentle pressure is applied backward to tilt the head. The fingertips of the other hand are positioned...
514