Related Experiment Video
Updated: Jan 13, 2026

Using Extraordinary Optical Transmission to Quantify Cardiac Biomarkers in Human Serum
Published on: December 13, 2017
Clinical Performance of a Novel Point-of-Care High-Sensitivity Cardiac Troponin I Assay
Alan H B Wu1, Simon A Mahler2, Stacey L House3
1Department of Laboratory Medicine, University of California, San Francisco, CA, United States.
Background:
High-sensitivity cardiac troponin (hs-cTn) assays performed in the central laboratory have long turnaround times. This study evaluates the clinical performance of a novel bedside point-of-care (POC) hs-cTnI assay (i-STAT® hs-TnI) to aid in the diagnosis of myocardial infarction (MI).
Methods:
A prospective multisite cohort study involving emergency department (ED) patients with symptoms suggestive of acute coronary syndrome was conducted at 28 sites across the United States. Serial whole blood samples were collected at 0-1 h and >1-3 h from ED presentation. Primary outcome was MI or non-MI. Sensitivity, specificity, and negative predictive value (NPV) for MI were calculated with 97.5% lower confidence limits (LCL) using the female, male, and overall 99th percentile upper reference limit (URL, 13, 28, and 21 ng/L, respectively). Post hoc analysis was also conducted for initial measurements below the limit of quantification (LoQ, 2.9 ng/L) and <5.0 ng/L.
Results:
The cohort included 3582 patients, median age 59 y, 64% female. MI incidence was 8.6%. Sensitivity and specificity at >1-3 h were 96.6% (LCL 91.7%) and 82.1% (80.2%) using the female URL; 90.7% (LCL 84.1%) and 83.9% (LCL 81.4%) using the male URL; and 92.4% (LCL 88.3%) and 85.6% (LCL 84.2%) using the overall URL. NPVs for MI of initial hs-cTnI measurements
Conclusions:
The whole blood POC i-STAT hs-TnI assay demonstrated performance characteristics similar to other hs-cTn assays and achieved high sensitivity and NPV for initial measurements
Clinicaltrials.Gov Registration Number:
NCT05629572.
More Related Videos
10:12A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay
Published on: August 8, 2013
08:58Fully Automated Centrifugal Microfluidic Device for Ultrasensitive Protein Detection from Whole Blood
Published on: April 16, 2016
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Acute Coronary Syndrome III: Diagnostic Studies
Acute Coronary Syndrome I: Introduction