Sound Source Localization Testing in Single-sided Deafness Following Bone Conduction Intervention
Published on: December 20, 2024
Maximum stable gain and fitting range of percutaneously applied bone conduction devices
Gertjan Dingemanse1, Martijn Toll1
1Department of Otorhinolaryngology and Head and Neck Surgery, Erasmus MC, University Medical Center Rotterdam, Rotterdam, The Netherlands.
Maximum stable gain (MSG) for bone conduction devices (BCDs) is limited by feedback, impacting speech intelligibility for mild-to-moderate hearing loss. The study determined an MSG-based fitting range and measured the aided dynamic range.
Area of Science:
- Audiology
- Biomedical Engineering
- Hearing Science
Background:
- Bone conduction devices (BCDs) offer an alternative for hearing loss.
- Understanding the gain limits and fitting ranges of BCDs is crucial for effective use.
Purpose of the Study:
- To measure the maximum stable gain (MSG) and its variability in users of percutaneous power BCDs.
- To establish an MSG-based fitting range for BCDs.
- To assess the available aided dynamic range with BCDs.
Main Methods:
- MSG was measured with the BCD on a participant's head and output on a skull simulator.
- A fitting range was determined using the speech intelligibility index.
- 42 adult BCD users participated in the study.
Main Results:
- MSG averaged 13.3 ± 3.2 dB for device 1 and 16.1 ± 5.9 dB for device 2.
- The fitting range for good speech understanding was 35-40 dB HL, and 45-50 dB HL for moderate-to-sufficient understanding.
- Output was generally not limited by maximum force, except at low frequencies.
Conclusions:
- The maximum usable gain in percutaneous power BCDs is limited by feedback, reducing speech intelligibility.
- BCDs provide sufficient speech intelligibility primarily for mild-to-moderate sensorineural hearing loss.
- The aided dynamic range is adequate for hearing thresholds within the determined fitting range.