Related Experiment Video
Updated: Jan 6, 2026

A Simplified Stepwise Approach to Echo Guidance during Percutaneous Mitral Valve Repair
Published on: October 16, 2021
Incidence, predictors, and clinical impact of hypoattenuating leaflet thickening following SAPIEN 3 Ultra RESILIA
Kenichi Ishizu1, Shinichi Shirai1, Masaomi Hayashi1
1Department of Cardiology, Kokura Memorial Hospital, Kitakyushu, Japan.
Background:
The latest-generation SAPIEN 3 Ultra RESILIA (S3UR) transcatheter heart valve (THV) incorporates several changes in leaflet design, including an improved anticalcification coating and modified commissural attachment. There are no established data on hypoattenuating leaflet thickening (HALT) following transcatheter aortic valve implantation (TAVI) using the S3UR.
Aims:
Our study aimed to elucidate the clinical features of HALT following S3UR implantation.
Methods:
As a subset of the OCEAN (Optimized CathEter vAlvular INtervention)-TAVI registry, we prospectively assessed patients who underwent cardiac computed tomography (CT) 30 days after S3UR implantation. HALT and potentially relevant THV geometry were analysed using four-dimensional CT data by an independent core laboratory.
Results:
Of the 445 patients studied, HALT was detected in 95 patients (21.3%) 30 days after TAVI. The modification of the commissural attachment specific to the 20 mm and 23 mm S3UR THVs did not affect the incidence of HALT (22.1% for ≤23 mm; 20.2% for ≥26 mm; p=0.636). The hourglass-shaped THV frame (p<0.001) and asymmetricity of THV leaflets (p=0.002) were independently associated with HALT development. A trend toward higher mean aortic gradients at 30 days with greater degrees of HALT (HALT >25% vs HALT ≤25%: 10.3 [interquartile range [IQR] 7.0-13.0] mmHg vs 8.6 [IQR 6.3-11.6] mmHg; p=0.007; HALT >50% vs HALT ≤50%: 11.5 [IQR 7.0-14.3] mmHg vs 8.9 [IQR 6.3-11.9] mmHg; p=0.002) was noted.
Conclusions:
The incidence of HALT for the S3UR was comparable with the already reported incidences for the previous-generation SAPIEN 3 THV. Given the haemodynamic impact of HALT severity and multiplicity, strategic planning to avoid deformation of the implanted THV might be required. (Clinical trial registration: UMIN000020423).