In vivoevaluation of human calf blood pressure by using different external compression strategies
Hanhao Liu1, Yawei Wang1, Shuai Tian2
1Key Laboratory for Biomechanics and Mechanobiology of Ministry of Education, Beijing Advanced Innovation Center for Biomedical Engineering, School of Biological Science and Medical Engineering, Beihang University, Beijing, People's Republic of China.
Abstract:
Objective.Calf blood pressure (CBP) plays an important role in various clinical applications, such as determining the appropriate cuff pressure for compression therapy and diagnosis of lower extremity vascular diseases, which necessitates the integration of built-in measurement methods within the device. This study aimed to investigate the differences in human CBP assessments resulting from the application of variousin vivoexternal compression strategies.Approach.An experimental procedure incorporating different compression strategies, specially the low-pressure-sustained mode, the slow-inflation (SI) mode and the slow-deflation (SD) mode, was conducted to capture the dynamic responses of the photoplethysmography (PPG) signal for CBP evaluation. Nineteen subjects, including 13 males and 6 females, participated in this experimental study. Feature points related to CBP were extracted from dynamic responses of the PPG signal and subjected to statistical analysis. A lumped parameter model of the human lower extremity was developed to assist in analyzing the biomechanical factors underlying these differences.Main results.The experimental results indicated that the dynamic behaviors of the PPG signal followed clear and consistent patterns, and the CBP value derived from the PPG signal in the SI mode was significantly higher than that obtained in the SD mode. Model simulation results showed that the differences in the evaluated values of the CBP between the SI and SD modes were caused by the different collapse processes of calf arteries and veins.Significance.This study could help understand the forced collapse process of blood vessels in the calf and inspire new ideas on CBP evaluation and personalized compression therapy.
Related Concept Videos
Assessing Blood pressure in the Leg
Preparation:
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.