Why and When ICD Leads Are Extracted: Does the ICD Lead Model Influence Lead Survival?
Andrzej Kutarski1, Wojciech Jacheć1, Paweł Stefańczyk2
12nd Department of Cardiology, Faculty of Medical Sciences in Zabrze, Medical University of Silesia, 40-055 Katowice, Poland.
Medicina (Kaunas, Lithuania)
|November 27, 2025
Summary
Implantable cardioverter-defibrillator (ICD) leads last three years less than pacemaker (PM) leads. Older, thicker ICD leads survive longer than newer, thinner models, with infection and mechanical damage being primary failure reasons.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Analysis
Background:
- Limited data exists on implantable cardioverter-defibrillator (ICD) lead longevity.
- Understanding ICD lead failure modes and model-specific lifespans is crucial for patient care.
Purpose of the Study:
- To compare the survival duration of different ICD lead models.
- To identify the primary reasons for ICD lead extraction.
Main Methods:
- Analysis of 3929 transvenous lead extractions (TLEs), including 1068 ICD lead extractions.
- Comparison of lead survival times and failure causes across various ICD lead models.
Main Results:
- The median ICD lead survival was 61 months, three years shorter than pacemaker (PM) leads.
- Common extraction reasons included infection (35.6%), mechanical damage (33.6%), and dysfunction (20.8%).
- Older, thicker ICD lead models (e.g., Sprint, Ventritex, Kainox) demonstrated significantly longer survival (nearly double) compared to thinner, modern leads (e.g., Riata, Sprint Fidelis).
Conclusions:
- ICD lead survival is notably shorter than PM leads.
- Infection and mechanical damage are the leading causes of ICD lead failure.
- Despite design advancements, the overall lifespan of ICD leads has not significantly improved, with older models outperforming newer ones in durability.