Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
11.7K
Normalization of temperature effects for quality assurance of quantitative prostate apparent diffusion coefficient
Ken-Pin Hwang1, Joshua Yung1, R Jason Stafford1
1Department of Imaging Physics, The University of Texas M.D. Anderson Cancer Center, Houston, Texas, USA.
Journal of Applied Clinical Medical Physics
|November 27, 2025
Summary
A new quality assurance procedure simplifies prostate MRI diffusion assessments by using a temperature model to ensure consistent Apparent Diffusion Coefficient (ADC) measurements across scanners without strict temperature control.
Area of Science:
- Magnetic Resonance Imaging
- Medical Physics
- Prostate Cancer Imaging
Background:
- Apparent Diffusion Coefficient (ADC) in diffusion-weighted imaging correlates with prostate tumor aggressiveness.
- Variability in ADC measurements across scanner networks necessitates quality assurance (QA) for prostate MRI protocols.
- Traditional ADC reproducibility assessments require strict 0°C phantom temperature control, posing a logistical challenge.
Purpose of the Study:
- To establish a room-temperature QA procedure for assessing the reproducibility of ADC measurements in prostate diffusion MRI.
- To develop a model correlating ADC, phantom gel concentration, and temperature for accurate normalization.
Main Methods:
- Diffusion phantom imaging across multiple 1.5T and 3T scanners at different locations.
- ADC measurements were taken at varying phantom gel concentrations.
- A temperature-dependent ADC model was applied to normalize measurements to a reference temperature (20°C).
Main Results:
- Temperature normalization reduced the variability (percent difference) in ADC measurements across scanners.
- The coefficient of variation (CV) for clinically relevant ADC values decreased from 5.0% to 3.4% post-normalization.
- The CV for pure water ADC decreased from 3.3% to 1.3% after normalization.
Conclusions:
- A novel QA procedure utilizing a temperature model eliminates the need for strict phantom temperature control.
- This method facilitates reliable ADC reproducibility assessments for prostate diffusion MRI across extensive scanner networks.
- The developed QA procedure enhances the consistency and accuracy of quantitative ADC measurements in multi-center studies.
Keywords:
apparent diffusion coefficientdiffusion weighted imagingprostate imagingquality assurancetemperature effects on diffusion