Clinical application of 3D-printed bolus in postmastectomy radiotherapy
Qianqian Liu1, Xuming Chen1, Zhekai Hu1
1Department of Radiation Oncology, Shanghai General Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, P. R. China.
Purpose:
To evaluate the fitting accuracy and dose influence of 3D-printed bolus (3DPB) during postmastectomy radiotherapy.
Methods:
Air gaps were assessed using cone-beam computed tomography (CBCT) images for 27 breast cancer patients who received postmastectomy radiotherapy. Then a CIRS thoracic phantom was used to evaluate the dosimetric effects of different air gap sizes.
Results:
In CBCT images of patients, compared with conventional bolus (CB), 3DPB significantly reduced both the maximum and total air gap volumes by an average between 103.04 cm3 and 124.63 cm3, respectively. The average maximum gap height was 0.53 ± 0.20 cm for 3DPB and 2.23 ± 0.41 cm for CB (p<0.05). In the results as obtained from the phantom simulation, as gap height increased from 0.5 cm to 3.5 cm, D95 (dose received by 95% of the target volume) and D98 (dose received by 98% of the target volume) values for the superficial skin zone (SSZ), clinical target volume (CTV), and planning target volume (PTV) all decreased insignificantly. When the gap exceeded 2 cm, dose reductions became more pronounced in SSZ. Similar trends were observed in the CTV and PTV.
Conclusions:
For breast cancer radiotherapy patients immobilized using a vacuum bag, as evidenced by the CBCT images and phantom simulation, 3DPB provides a superior anatomical fit, reducing air gaps and associated dose loss in postmastectomy radiotherapy. When the air gap height ≤ 2 cm, dose attenuation with CB remains clinically acceptable, suggesting that 3D printing may not be necessary unless the skin in the gap area is at high risk for recurrence. However, for gaps > 2 cm, the dose reduction becomes significant, warranting the use of 3D-printed boluses to ensure adequate target coverage.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Peptic Ulcer Disease V: Surgical Management and Nursing Care
Surgical Interventions for Peptic Ulcer Disease
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Urinary Tract Infection III: Diagnostic Studies and Interprofessional Care
Imaging Studies II: Ultrasonography
Imaging Studies VI: Voiding Cystourethrography and Cystography