Joint range-modulator and spot optimization for Bragg-peak proton FLASH radiotherapy
Jiayue Han1, Aoxiang Wang2, Ya-Nan Zhu3
1Department of Radiation Oncology, University of Kansas Medical Center, Kansas City, Kansas, USA.
Medical Physics
|November 29, 2025
Summary
A new Joint Range-Modulator and Spot Optimization (JRSO) method improves proton FLASH (pFLASH) radiotherapy planning by simultaneously optimizing beam parameters. This novel approach enhances plan quality and expands the clinical applicability of pFLASH therapy.
Area of Science:
- Medical Physics
- Radiation Oncology
- Computational Biology
Background:
- Ultra-high-dose-rate (UHDR) radiation therapy, particularly proton FLASH (pFLASH), shows potential for reduced organ-at-risk (OAR) toxicity.
- Current pFLASH treatment planning involves a two-step process: intensity-modulated proton therapy (IMPT) plan generation followed by patient-specific range modulator (PSRM) optimization.
- The interplay between spot weights and PSRM design in pFLASH therapy is not fully understood, potentially limiting dosimetric and radiobiological benefits.
Purpose of the Study:
- To introduce a novel alternating optimization framework, Joint Range-Modulator and Spot Optimization (JRSO), for simultaneous optimization of PSRM and spot weights.
- To enhance the plan quality of conformal pFLASH radiotherapy through integrated optimization.
Main Methods:
- JRSO framework developed for simultaneous optimization, removing the need for a one-to-one correspondence between beam spots and PSRM pins.
- Proton beam diffusion modeled using a Gaussian function.
- Iterative alternating updates of PSRM design and spot weights, starting from an IMPT-derived solution, while adhering to constraints like minimum monitor units (MMU).
Main Results:
- JRSO demonstrated superior plan quality compared to conventional methods.
- In a head-and-neck (HN) case, JRSO reduced the objective function value (0.46 to 0.26), lowered maximum target dose (117.6% to 107.1%), improved conformity index (0.74 to 0.87), and decreased ROI effective dose (6.50 Gy to 6.10 Gy).
Conclusions:
- The proposed JRSO method offers an advancement for conformal pFLASH radiotherapy planning.
- JRSO outperforms the conventional approach and may broaden PSRM applicability in complex clinical scenarios, including spot-pin misalignments.
- Numerical results confirm the robustness and efficiency of the JRSO method.
More Related Videos
Related Concept Videos
Radiation: Applications
1.7K
The average temperature of Earth is the subject of much current discussion. Earth is in radiative contact with both the Sun and dark space; it receives almost all its energy from the radiation of the Sun and reflects some of it into outer space. Dark space is very cold, about 3 K, so Earth radiates energy into it. For instance, heat transfer occurs from soil and grasses, the rate of which can be so rapid that frost can occur on clear summer evenings, even in warm latitudes.
The average...
The average...
1.7K
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
1.7K
A pulse is a short burst of radio waves distributed over a range of frequencies that simultaneously excites all the nuclei in the sample. Upon passing a radio frequency pulse along the x-axis, the nuclei absorb energy corresponding to their Larmor frequencies and achieve resonance. This shifts the net magnetization vector from the z-axis toward the transverse plane. This angle of rotation of the magnetization vector, or the flip angle, is proportional to the duration and intensity of the pulse.
1.7K
Positron Emission Tomography
6.8K
Positron emission tomography (PET) is a medical imaging technique involving radiopharmaceuticals — substances that emit short-lived radiation. Although the first PET scanner was introduced in 1961, it took 15 more years before radiopharmaceuticals were combined with the technique and revolutionized its potential.
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
6.8K