Evaluation of Refractive Measurements Using a Handheld Autorefractor (Retinomax K+ Screen) Versus Streak Retinoscopy
Juliana Giselbrecht1, Barbara Nell1
1Department of Companion Animals and Horses, University of Veterinary Medicine Vienna, Vienna, Austria.
Veterinary Ophthalmology
|December 1, 2025
Summary
The Retinomax K+ Screen handheld autorefractor overestimates hyperopia in dogs compared to manual streak retinoscopy. Its significant variability and reduced reliability question its clinical precision as a screening tool.
Area of Science:
- Veterinary ophthalmology
- Ophthalmic diagnostics
- Refractive error assessment
Background:
- Accurate refractive assessment is crucial for diagnosing ocular conditions in dogs.
- Handheld autorefractors offer potential for rapid, non-invasive measurements.
- Evaluating new diagnostic tools is essential for improving veterinary eye care.
Purpose of the Study:
- To compare the accuracy of the Retinomax K+ Screen handheld autorefractor with manual streak retinoscopy in non-cyclopleged dogs.
- To assess the clinical applicability and reliability of the autorefractor for canine refractive error evaluation.
Main Methods:
- Thirty-four dogs (68 eyes) underwent ophthalmic examination and refractive assessment.
- Measurements were obtained using both streak retinoscopy and the Retinomax K+ Screen handheld autorefractor.
- Statistical analyses included Wilcoxon signed-rank test, Bland-Altman analysis, and Spearman's rank correlation.
Main Results:
- Autorefraction was successful in 57 of 68 eyes; 11 were excluded due to poor quality.
- Autorefraction systematically overestimated hyperopia compared to retinoscopy (0.98 D vs 0.13 D).
- Wide limits of agreement (-2.52 to 4.10 D) and moderate correlation (ρ=0.49) indicated substantial variability.
Conclusions:
- The Retinomax K+ Screen provides rapid measurements but overestimates hyperopia in dogs.
- Considerable variability and reduced reliability limit its clinical precision.
- Its use as a screening tool in clinical settings is questionable due to inconsistent results.