Related Experiment Video
Updated: Jan 9, 2026

03:07
Single-Port Robotic-assisted Transaxillary Breast-conserving Surgery: A Prospective, Single-arm, Non-randomized Phase IIa Clinical Trial
Published on: August 19, 2025
772
Improved surgical accuracy in total knee arthroplasty using the ROSA® knee system: A randomised-controlled unblinded
Alexander Möller1, Feras Kasabji1, Maximilian Fischer1
1Center for Orthopedics, Trauma Surgery and Rehabilitation Medicine, University Medicine Greifswald, Greifswald, Germany.
Summary
Robotic-assisted total knee arthroplasty (RATKA) using the ROSA® system improved limb alignment compared to conventional TKA. Patient outcomes were similar in the short term, with further research needed on long-term implant survival.
Area of Science:
- Orthopedic surgery
- Robotics in medicine
- Biomechanical engineering
Background:
- Robotic-assisted total knee arthroplasty (RATKA) shows potential for enhanced surgical precision and implant alignment over conventional TKA (CTKA).
- Limited randomized-controlled trial (RCT) evidence exists, and various robotic systems are available.
- This study compares the ROSA® knee system for RATKA against CTKA.
Purpose of the Study:
- To prospectively compare surgical and patient-reported outcomes of RATKA using the ROSA® knee system versus CTKA.
- To evaluate implant alignment accuracy and patient satisfaction.
- To assess secondary outcomes including range of motion and length of stay.
Main Methods:
- A monocentric, unblinded RCT included 77 patients undergoing TKA with a minimum 1-year follow-up.
- Primary endpoints: postoperative implant alignment, Oxford knee score (OKS), and EQ-5D-5L quality of life scores.
- Secondary endpoints: range of motion, blood loss, length of stay, return to work, and physiotherapy needs.
Main Results:
- RATKA demonstrated significantly improved limb alignment, with a lower frequency of extreme malalignment compared to CTKA (18% vs. 38%, p=0.047).
- Postoperative improvements in OKS and EQ-5D-5L scores were comparable between groups (p>0.05).
- Preoperative OKS and cruciate-retaining implant bearing were significant predictors of improved outcomes.
Conclusions:
- RATKA with the ROSA® knee system enhances implant positioning and radiological alignment compared to CTKA.
- Short-term clinical outcomes were similar between the two methods.
- Long-term RCT data are necessary to ascertain if improved RATKA accuracy translates to superior implant survival and reduced revision rates.