Optimization and Reliability Assessment of Ultrasound Shear Wave Elastography of the Rectus Femoris: A Methodological
Yating Wang1, Xiaona Liu1, Xuemei Liu1
1Department of Ultrasound, Beijing Hospital of Traditional Chinese Medicine, Capital Medical University, Beijing, China.
Objective:
To evaluate effects of various sections, operational methods, and measurement techniques of shear wave elastography (SWE) on the assessment of rectus femoris stiffness.
Methods:
This study included 43 healthy volunteers who underwent rectus femoris SWE using different sections (cross-sectional and longitudinal), operation methods (hand-held, thick coating couplant, and acoustic pad), and measurement methods (circular, multipoint, and tracing). Two sonographers evaluated the inter-observer and intra-observer consistency of SWE longitudinal-section measurements for the first 20 healthy volunteers.
Results:
The longitudinal section had lower median values of shear wave velocity (SWV) and shear modulus (1.7 m/s vs. 1.9 m/s, 9.3 kPa vs. 11.4 kPa), whereas the transverse section exhibited higher coefficients of variation (17.14% vs. 10.94%, 33.80% vs. 27.70%). Excellent inter-observer agreement on SWV and shear modulus was observed for the longitudinal section (ICC: 0.97, 0.96, 95% CI: 0.883-0.991, 0.800-0.988; p < 0.001). The intra-observer agreement for SWV was substantial (ICC = 0.64) while that for the shear modulus was moderate (ICC = 0.23). Median SWVs of the handheld, thick-coated couplant, and acoustic-pad methods decreased in order (1.9 m/s, 1.7 m/s, 1.6 m/s, respectively). The handheld method had the lowest coefficient of variation (CV) compared to the other two methods (10.94%, 13.87%, and 14.56%, respectively). No statistically significant differences were observed in SWV measurements and CV between circle, multipoint, and tracing methods (p > 0.05).
Conclusion:
The longitudinal section and handheld method offered enhanced stability when measuring muscle stiffness; shear SWV exhibited superior consistency compared to shear modulus. The region-of-interest tracing method was straightforward and had an extensive measurement range.