Related Experiment Video
Updated: Jan 8, 2026

Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Transcatheter Implantation of Balloon-Expandable Valve in a Repaired Native Tricuspid Valve Without Annuloplasty Ring
Michael Yang1, Marissa A Munsayac1, Steven C Romero1
1Division of Cardiology, Scripps Clinic, La Jolla, California, USA.
Abstract:
Late complications of repaired congenital heart disease are increasingly common, including tricuspid stenosis and regurgitation. We present a case of a 36-year-old female with tetralogy of Fallot, 1.5 ventricle repair, and prior tricuspid repair using homograft tissue without an annuloplasty ring, who developed worsening congestive heart failure. Transesophageal echocardiogram revealed severe tricuspid stenosis and regurgitation. We performed transcatheter tricuspid valve replacement with a SAPIEN S3 Ultra RESILIA valve within a CP-covered stent. To our knowledge, this is the first implantation of the SAPIEN S3 Ultra RESILIA valve in the tricuspid position without an annuloplasty ring, and the first implantation of a balloon-expandable valve in a native tricuspid valve without extracorporeal membrane oxygenation support.