Simulation Study of a Hyperbolic Linear Ion Trap with Asymmetric Curvature Radii
Yang Li1,2, Yuanjiang Luo1,2, Lingwen Kong1,2
1Anhui Province Key Laboratory of Medical Physics and Technology, Hefei Cancer Hospital of CAS, Institute of Health and Medical Technology, Hefei Institutes of Physical Science, Chinese Academy of Sciences, 230031 Hefei, P. R. China.
None:
Linear ion traps (LITs) with simplified geometries have shown potential for miniaturized mass spectrometry systems, though structural simplifications often introduce nonlinear higher-order fields that compromise performance. This study investigates asymmetric hyperbolic electrode modifications to address the inherent 50% detection efficiency limitation in miniaturized LITs caused by bidirectional ion ejection. The research employed numerical simulations to evaluate a geometrically modified LIT design. Electrode asymmetry was introduced through independent adjustments of vertex curvature radii and bending angles in opposing x-direction electrodes, creating an asymmetric trapping field. Structural optimization involved initial x-axis elongation of a centrosymmetric ion trap followed by controlled geometric modifications. Field analysis confirmed the incorporation of odd-order multipole components through these adjustments. Experimental results demonstrated approximately 90% ion ejection efficiency from a single trap side when combining higher-order odd and even multipole fields. This work provides valuable insights into the design of asymmetric ion traps, offering a practical solution for high-performance portable MS development.