Evaluation of an event-driven 3FI ASIC for spectroscopic X-ray detection with synchrotron radiation
Piotr Maj1, Grzegorz W Deptuch1, Dominik S Gorni1
1Instrumentation Department, Brookhaven National Laboratory, PO Box 5000, Upton, NY 11973-5000, USA.
Abstract:
The novel design and evaluation on the NSLS-II beamline of the 3FI application-specific integrated circuit (ASIC) bump-bonded to a simple, planar, 2D segmented silicon sensor are presented. The ASIC was developed for full-field fluorescence spectral X-ray imaging (3FI). It is a small-scale prototype that features a square array of 32 × 32 pixels, and the size of the pixels is 100 µm × 100 µm. The ASIC was implemented in a 65 nm CMOS integrated circuit fabrication process. Each pixel incorporates a charge-sensitive amplifier, a shaping filter, a discriminator, a peak detector and a sample-and-hold circuit, allowing detection of events and storage of signal amplitudes. The system operates in a frameless event-driven readout mode, outputting analog values for threshold-triggered events, allowing high-speed multi-element X-ray fluorescence data acquisition. The 3FI ASIC achieves per-channel spectrometric performance at a power consumption of only 200 µW per pixel, with nearly all dissipation confined to the analog front-end. An energy resolution is measured at the level of 308 eV full width at half-maximum (FWHM) at 8.04 keV (Cu Kα), and 138 eV FWHM at 3.69 keV (Ca Kα). This per-pixel capability makes the prototype suitable for in situ trace element microanalysis in biological and environmental studies. Moreover, the frameless architecture of the detector is designed to address limitations of conventional X-ray fluorescence microscopy, which typically requires mechanical scanning, by enabling continuous high-throughput data acquisition in future full-field implementations.
Related Concept Videos
X-ray Imaging
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....