Gaia: segmented germanium detector for high-energy X-ray fluorescence and spectroscopic imaging
Abdul K Rumaiz1, Francesca Capocasa2, Anthony J Kuczewski1
1NSLS II, Brookhaven National Laboratory, Upton, NY 11973, USA.
None:
We present Gaia, a monolithic array of 96 high-purity germanium pixel detectors integrated with a custom low-noise application-specific integrated circuit (ASIC) and a field-programmable gate array (FPGA)-based data acquisition system. The sensor operates at ∼100 K using a commercial closed-cycle cryocooler, with the in-vacuum electronics thermally isolated from the cold finger to ensure thermal stability. The system demonstrates an average energy resolution of 711 eV at 122 keV, measured using a 57Co source, and 253 eV at 5.89 keV, measured with 55Fe across all channels. The readout architecture incorporates a high-performance FPGA paired with a dual-core ARM processor, forming a complete embedded Linux-based computing platform. Communication between the processor and FPGA is handled via memory-mapped I/O, and data are streamed over high-speed gigabit Ethernet. A full-scale 384-pixel Gaia detector, based on this 96-element module, is currently under fabrication.