Phase Contrast Magnetic Resonance Imaging in the Rat Common Carotid Artery
Published on: September 5, 2018
Arterial flow volume measurement using signal intensity gradient versus phase contrast
Chan-Hyuk Lee1, Sang Hyuk Lee2, Hyo-Sung Kwak3
1Department of Neurology, Ulsan University Hospital, University of Ulsan College of Medicine, Ulsan, Republic of Korea.
Signal intensity gradient (SIG) volume from Time-of-flight magnetic resonance angiography (TOF-MRA) accurately measures cerebral blood flow. This method shows high correlation with two-dimensional phase-contrast MR (PC-MR) in both experimental and human studies, aiding cerebrovascular disease understanding.
Area of Science:
- Medical Imaging
- Cardiovascular Research
- Neuroimaging
Background:
- Accurate measurement of cerebral blood flow is vital for diagnosing and managing cerebrovascular diseases.
- Time-of-flight magnetic resonance angiography (TOF-MRA) is a common technique, but its ability to quantify flow volume needs validation.
- Signal intensity gradient (SIG) offers a potential method for flow volume calculation within TOF-MRA.
Purpose of the Study:
- To compare blood flow volumes measured by SIG in TOF-MRA with those obtained by two-dimensional phase-contrast MR (PC-MR).
- To assess the correlation and agreement between SIG volume and PC-MR flow volumes in experimental setups and human subjects.
- To evaluate the clinical applicability of SIG volume from TOF-MRA for assessing cerebral blood flow.
Main Methods:
- Experimental study using four varying-sized tubes to measure flow volumes with both PC-MR and TOF-MRA.
- Prospective cross-sectional study involving healthy human participants, examining internal carotid and vertebral arteries.
- Statistical analysis included Pearson's correlation coefficients, coefficients of determination (R²), and Bland-Altman plots to assess agreement.
Main Results:
- In tubal experiments (n=80), flow volumes showed a high correlation (r=0.98, R²=0.96, p<0.001) between TOF-MRA SIG volume and PC-MR.
- In human studies, 50 internal carotid and 49 vertebral arteries were analyzed.
- Cerebral artery flow volumes demonstrated a strong correlation (r=0.93, R²=0.87, p<0.001) between the two methods.
Conclusions:
- Calculated SIG volumes from TOF-MRA are highly correlated with blood flow volumes measured by PC-MR.
- The SIG method provides a reliable approach for quantifying blood flow in extracranial cerebral arteries.
- This finding supports the use of TOF-MRA with SIG calculation for improved cerebrovascular assessment.
Related Concept Videos
Blood Flow
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure
Gradually Varying Flow