Early Safety and Feasibility Study of a Novel Transcatheter Aortic Valve With Polymeric Leaflets
Xiaochun Zhang1, Qinchun Jin1, Wenzhi Pan1
1Department of Cardiology, Zhongshan Hospital, Fudan University, Shanghai Institute of Cardiovascular Diseases, Shanghai, China; State Key Laboratory of Cardiovascular Diseases, Zhongshan Hospital, Fudan University, Shanghai, China; NHC Key Laboratory of Ischemic Heart Diseases Fudan University, Shanghai, China; Key Laboratory of Viral Heart Diseases, Chinese Academy of Medical Sciences, Shanghai, China; National Clinical Research Center for Interventional Medicine, Shanghai, China.
Background:
Transcatheter aortic valve replacement is a promising alternative treatment for patients with severe aortic stenosis (AS), but uncertainties remain regarding its long-term durability.
Objectives:
This study aimed to evaluate the safety and feasibility of Sikelia transcatheter heart valve with polymeric leaflets in AS.
Methods:
In this first-in-human, prospective, nonrandomized, single-arm, multicenter study, patients with severe symptomatic AS eligible for the novel valve were recruited. Clinical outcomes and hemodynamic performance were reported at discharge and 30-day, 6-month, and 1-year follow-up.
Results:
Twelve patients (73.75 ± 7.28 years, 25% female, mean Society of Thoracic Surgeons score 4.51% [4.07%-4.87%], 66.67% bicuspid valve) were enrolled. Procedural success was achieved in 83.33% (1 fully device retrieval and 1 death due to coronary obstruction). All patients successfully underwent transcatheter aortic valve replacement were in NYHA functional class I/II at the 30-day follow-up, and significant improvement was shown in 6-minute walk distance and Kansas City Cardiomyopathy Questionnaire score (all P < 0.001). At 1 year, the effective orifice area (2.31 ± 0.33 cm2 vs 0.82 ± 0.14 cm2) and mean pressure gradient (6.33 ± 2.54 mm Hg vs 47.50 ± 13.82 mm Hg) improved substantially, so did those in bicuspid patients (all P < 0.001). No moderate/severe paravalvular leakage was recorded, and computed tomography scans revealed no evidence of subclinical leaflet thrombosis during follow-up. After discharge, no patients reported death, stroke, new pacemakers, or valve-related reintervention.
Conclusions:
The preliminary outcomes of the novel transcatheter heart valve demonstrated a favorable hemodynamic performance with an acceptable safety profile at 1 year. Long-term follow-up is warranted to assess its durability in treating patients with AS.
Related Concept Videos
Survey Safety
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Household Wiring And Electrical Safety
Olefin Metathesis Polymerization: Ring-Opening Metathesis Polymerization (ROMP)
Actin Polymerization
The nucleation phase involves forming a stable nucleus consisting of three actin monomers to form a new actin filament. Actin-binding proteins such as formins and Arp2/3 complex help filament growth post-nucleation. The Formins form straight...
Mitral Valve Prolapse I: Introduction