Comparison of a new non-contact automatic ophthalmic optical biometer with an SS-OCT based optical biometer ARGOS
Xuanqiao Lin1, Jiawei Li1, Yue Li1
1Eye Institute and Department of Ophthalmology, Institute for Medical and Engineering Innovation, Eye & ENT Hospital, Fudan University; NHC Key Laboratory of Myopia and Related Eye Diseases (Fudan University), Key Laboratory of Myopia and Related Eye Diseases, Chinese Academy of Medical Sciences, Shanghai, China; Shanghai Research Center of Ophthalmology and Optometry, Shanghai, China.
Purpose:
To evaluate the agreement of ocular biometric parameters measured by a novel all-in-one optical low coherence reflectometry (OLCR) biometer (SW-9000 Plus) compared with a swept-source optical coherence tomography (SS-OCT) biometer (ARGOS).
Methods:
48 eyes of 48 healthy subjects were included. The following parameters were obtained: axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), mean keratometry (Km), astigmatism (AST), J0, J45, corneal diameter (CD) and pupil diameter (PD). Bland-Altman plots and 95% limits of agreement (LoA) were utilized to confirm the concordance between the devices. Participants were categorized into two groups based on AL to facilitate a subgroup analysis (AL > 25 mm or AL < 25 mm).
Results:
The new biometer showed small differences with SS-OCT: AL (0.01 ± 0.03 mm), CCT (4.65 ± 4.29 μm), ACD (-0.11 ± 0.05 mm), LT (0.03 ± 0.08 mm), Km (-0.08 ± 0.18 D), AST (-0.03 ± 0.17 D), J0 (0 ± 0.09 D), J45 (-0.05 ± 0.09 D), CD (-1.21 ± 0.36 mm), and PD (-0.37 ± 0.64 mm). For AL > 25 mm and AL < 25 mm, differences were 0.02 ± 0.03 mm and -0.01 ± 0.03 mm. Bland-Altman plots showed narrow 95% LoAs for all except PD (-1.62 to 0.87 mm) and CD (-1.91 to -0.51 mm).
Conclusions:
The novel all-in-one OLCR biometer showed high agreement with the SS-OCT device for most ocular biometry parameters in healthy eyes. CD and PD demonstrated lower agreement and should not be considered interchangeable between devices.
Related Concept Videos
Properties of Enantiomers and Optical Activity
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Automatic Processing and Automatic Social Behavior
Ophthalmic Drug Delivery Systems
Transmission-based Precautions I: Contact, Enteric, and Droplets
Contact Precautions:
Contact precautions are the measures taken to prevent the transmission of infectious agents, especially epidemiologically important microorganisms such as MRSA or influenza, primarily transmitted through direct or indirect contact with an...
The Sense of Self: Reflected Self-Appraisal and Social Comparison