Virucidal efficacy of a UV-C disinfection system for semi-critical medical devices according to EN 14476
Stefan Alexander Rudhart1, Klaus Wolfgang Georg Eichhorn1, Marvin Rausch2
1Department of Otorhinolaryngology, University Medical Center Bonn (UKB), 53127 Bonn, Germany.
Purpose:
Semi-critical medical devices, such as flexible endoscopes in otorhinolaryngology (ORL), require high-level disinfection due to their contact with mucous membranes. UV-C systems offer a rapid, non-toxic alternative to traditional reprocessing. This study evaluates the virucidal efficacy of the D60 UV-C disinfection system according to EN 14476 standards.
Materials And Methods:
Standardized tests were performed using frosted glass and thermoplastic elastomer (TPE) carriers inoculated with four indicator viruses: murine norovirus (MNV), poliovirus type 1, adenovirus type 5, and simian virus 777 (SV40). After drying, carriers underwent a 60 second UV-C cycle in the D60 system. Viral recovery and titration followed EN 14476 protocols, with infectivity measured by TCID50 assay.
Results:
All viruses showed significant reductions following UV-C treatment. Residual titers fell below the detection limit (3.2 lg TCID50/mL), corresponding to >4.0 lg reduction across all tested surfaces and virus types. Adenovirus showed the highest reduction (>4.5 lg).
Conclusion:
The D60 system achieved broad-spectrum virucidal efficacy within one minute, meeting EN 14476 requirements. These findings support its use for safe, rapid reprocessing of semi-critical devices in high-turnover clinical settings, with potential to improve infection control and workflow efficiency.
Related Concept Videos
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Hand hygiene
Hand washing...
Chemical Agents for Microbial Control