Transition from passive scattering to pencil beam scanning: Physics commissioning of a proton radiosurgery system
Juliane Daartz1,2, Joost M Verburg1,2, Nicolas Depauw1,2
1Department of Radiation Oncology, Massachusetts General Hospital, Boston, MA, USA.
Background:
Our institution has provided proton stereotactic radiosurgery (PSRS) since 1961 using passive scattering (PS). A recent transition to pencil beam scanning with a new treatment planning system required extensive validation, particularly for small field applications.
Purpose:
This study details the clinical adaptation of our PSRS program to pencil beam scanning (PBS), emphasizing small field dosimetry and treatment delivery accuracy.Methods: Patient alignment via implanted fiducials was validated using a head phantom. Test fields were created to span clinically relevant parameters including distal target depth (3.4-17.9 cm), target thickness (1.0-3.0 cm), aperture diameter (0.9-3.0 cm), and aperture aspect ratios (1.0-3.0). Dosimetric validation of the in-house, vendor co-developed planning system included: (1) central axis depth-dose measurements in water with a micro-diamond detector; (2) absolute point dose measurements in solid water; and (3) two-dimensional relative dose distributions with Gafchromic film. Validation was supplemented by a commercial high-resolution ion chamber array.
Results:
End-to-end alignment using a hidden target approach showed a mean deviation of 0.6 ± 0.3 mm. Agreement with the treatment planning system (TPS) for 90% distal and proximal central axis depths was within 0.1 ± 0.7 mm and 0.7 ± 0.5 mm, respectively. Absolute isocenter dose agreement in water was within 1.6%. Gamma analysis (2%/1 mm, 10% threshold) yielded a 97.6% average pass rate. A patient-specific QA protocol for small fields was established and validated.
Conclusions:
We successfully transitioned our PSRS program to PBS, achieving high geometric and dosimetric accuracy suitable for small field proton radiosurgery.