Motion-Acuity Test for Visual Field Acuity Measurement with Motion-Defined Shapes
Published on: February 23, 2024
A Comparative Evaluation of 2 Rapid Strategies for Visual Field Testing: SITA Faster and SPARK Quick
Xiaoyan Wang1, Xiaoqian Bian2, Yize Chen2
1Ningbo Eye Hospital, Wenzhou Medical University, Ningbo, Zhejiang, China.
Ophthalmology. Glaucoma
|May 12, 2026
Summary
SITA Faster (SF) and SPARK Quick (SQ) significantly reduce visual field test times. SF offers comparable diagnostic accuracy to SITA Standard (SS), while SQ may underestimate visual field defects.
Area of Science:
- Ophthalmology
- Optometry
- Medical Diagnostics
Background:
- Visual field testing is crucial for glaucoma diagnosis and management.
- SITA Standard (SS) is the gold standard but can be time-consuming.
- Rapid visual field testing strategies aim to improve efficiency without compromising accuracy.
Purpose of the Study:
- To compare the diagnostic performance of SITA Faster (SF) and SPARK Quick (SQ) against SITA Standard (SS).
- To evaluate measurement accuracy, agreement, and repeatability of SF and SQ.
- To assess the impact of rapid testing strategies on clinical decision-making.
Main Methods:
- Prospective, cross-sectional study involving 118 eyes of 118 participants.
- Participants underwent SITA Standard (SS) 24-2, followed by randomly ordered SITA Faster (SF) and SPARK Quick (SQ) tests.
- Statistical analyses included ANOVA, ICC, Bland-Altman plots, and ROC curves to assess accuracy, agreement, and repeatability.
Main Results:
- Both SF and SQ significantly reduced test duration compared to SS (57.8% and 73.8% reduction, respectively).
- SF showed no significant difference in Mean Deviation (MD) and Pattern Standard Deviation (PSD) compared to SS.
- SQ demonstrated significantly better MD and PSD but had lower repeatability (ICC) and sensitivity for detecting defects (40.5%) compared to SF (83.8%).
Conclusions:
- SITA Faster (SF) provides comparable diagnostic performance to SITA Standard (SS) with significantly reduced testing time.
- SPARK Quick (SQ) also shortens test duration but may underestimate visual field defects and exhibits lower repeatability.
- SF appears to be a reliable alternative for rapid visual field testing in glaucoma patients.