In Vitro Assessment of 3T MRI Effects on Non-3T-Compatible Cardiac Implantable Electronic Devices
Yu Fukuoka1, Yasuo Sakurai2, Risa Kanai3
1Clinical Engineering Department, Saitama Medical University International Medical Center, Hidaka, Saitama, Japan.
Most cardiac implantable electronic devices (CIEDs) are safe for 3-Tesla (3T) MRI scans, though some pacemakers nearing elective replacement indication (ERI) may experience telemetry issues. ICDs generally showed no malfunctions during 3T MRI.
Area of Science:
- Cardiology
- Medical Imaging
- Biomedical Engineering
Background:
- The safety of cardiac implantable electronic devices (CIEDs) during 3-Tesla (3T) Magnetic Resonance Imaging (MRI) is not fully established, especially for devices not specifically designed for 3T environments.
- This study experimentally assessed the effects of 3T MRI on both 3T-compatible and non-3T-compatible CIEDs.
Purpose of the Study:
- To experimentally evaluate the safety and performance of various CIEDs when exposed to 3T MRI.
- To identify potential risks and malfunctions associated with 3T MRI in pacemakers and implantable cardioverter-defibrillators (ICDs).
Main Methods:
- Fifty-eight CIEDs (37 pacemakers, 21 ICDs) categorized as 3T MRI-conditional, 1.5-T MRI-conditional, or non-MRI-conditional were tested in a 3T MRI scanner.
- Devices were immersed in saline, connected to leads, and programmed to specific pacing modes with MRI mode ON or OFF.
- Interrogation before and after scanning assessed telemetry integrity, parameter retention, and pacing mode changes.
Main Results:
- The majority of CIEDs completed 3T MRI without significant malfunction or telemetry loss.
- Pacemakers, particularly those with low battery status nearing elective replacement indication (ERI), showed a higher incidence of telemetry failure or parameter reset.
- No reset or telemetry failure was observed in implantable cardioverter-defibrillators (ICDs).
Conclusions:
- While most CIEDs are safe for 3T MRI, pacemakers nearing ERI may present an increased risk of telemetry failure or reset.
- Non-negligible mode switching was observed in pacemakers with the MRI mode turned off.
- Overall, severe electrical malfunctions requiring device replacement were rare, and ICDs demonstrated robust safety during 3T MRI exposure.