S-ICD Oversensing of Leadless Atrial Pacing Spikes
Julia Ossi1, Mikhael F El-Chami2
1Department of Medicine, J. Willis Hurst Internal Medicine Residency Training Program, Emory University School of Medicine, Atlanta, Georgia, USA. Electronic address: https://twitter.com/Julia_Ossi.
Subcutaneous implantable cardioverter-defibrillators (S-ICDs) combined with atrial leadless pacemakers (LPs) can cause sensing interference. Adjusting the LP amplitude can resolve S-ICD oversensing issues.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Devices
Background:
- Subcutaneous implantable cardioverter-defibrillators (S-ICDs) and leadless pacemakers (LPs) are increasingly utilized.
- Limited data exist on the combined use of S-ICDs and LPs, particularly atrial LPs.