Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences01:20

Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences

Inductively coupled plasma–mass spectrometry (ICP–MS) is a highly selective and sensitive technique for accurate elemental analysis. Though the analysis of ICP–MS mass spectra is comparatively straightforward, it is affected by spectroscopic and non-spectroscopic interferences. Spectroscopic interferences arise when the plasma contains ionic species with an m/z value the same as the analyte ion. Spectroscopic interference can be categorized as isobaric, polyatomic ions, and refractory oxide ion...
Special considerations while measuring blood pressure01:28

Special considerations while measuring blood pressure

When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of apical radial pulse01:25

Assessment of apical radial pulse

Apical-Radial (A-R) Pulse Assessment
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Errors occurring during blood pressure monitoring01:25

Errors occurring during blood pressure monitoring

Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure01:10

Pre-Procedural Guidelines for Assessing Blood Pressure

Accurate blood pressure assessment is crucial for diagnosing and managing various health conditions. To ensure the reliability of these measurements, healthcare professionals must adhere to standardized pre-procedural guidelines. These guidelines enhance patient safety and improve the overall quality of healthcare. The following steps are essential for obtaining accurate and consistent blood pressure readings, from using the appropriate tools to ensuring effective communication with the patient.
Imaging Studies for Cardiovascular System IV: CMRI01:21

Imaging Studies for Cardiovascular System IV: CMRI

Cardiovascular magnetic resonance imaging, or CMRI, is a non-invasive diagnostic test that employs a magnetic field and radiofrequency waves to create precise images of the heart and arteries. It provides comprehensive information about cardiac anatomy, function, perfusion, and tissue characterization without ionizing radiation.IndicationsCMRI diagnoses various heart conditions, including tissue damage from heart attacks, ischemic heart disease, myocarditis, aortic issues (tears, aneurysms,...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

[Interference and misleading blood test results].

Ugeskrift for laeger·2026
Same author

Reshaping the progranulin/sortilin interaction for targeted degradation of extracellular proteins.

Cell chemical biology·2026
Same author

GFAP/UCH-L1 Assay for mTBI: Preanalytical Factors and Levels in Healthy Individuals-A Study on Danish Adults.

The journal of applied laboratory medicine·2026
Same author

[Porphyria cutanea tarda].

Ugeskrift for laeger·2025
Same author

[Porphyria cutanea tarda and hepatitis C infection].

Ugeskrift for laeger·2025
Same author

A low frequency damaging SORCS2 variant identified in a family with ADHD compromises receptor stability and quenches activity.

Molecular psychiatry·2025
Same journal

[Mosaicism as a cause of Cowden syndrome].

Ugeskrift for laeger·2026
Same journal

[Thyrotoxic crisis with cardiogenic shock].

Ugeskrift for laeger·2026
Same journal

[Mindfulness-based cognitive therapy in the treatment of depression].

Ugeskrift for laeger·2026
Same journal

[Fitness].

Ugeskrift for laeger·2026
Same journal

[Risk of developing cataract related to fluoroscopy].

Ugeskrift for laeger·2026
Same journal

[Obturator hernia with small bowel ileus presenting as hip pain in an elderly woman].

Ugeskrift for laeger·2026
See all related articles

Related Experiment Video

Updated: Jul 4, 2026

Phosphorus-31 Magnetic Resonance Spectroscopy: A Tool for Measuring In Vivo Mitochondrial Oxidative Phosphorylation Capacity in Human Skeletal Muscle
09:40

Phosphorus-31 Magnetic Resonance Spectroscopy: A Tool for Measuring In Vivo Mitochondrial Oxidative Phosphorylation Capacity in Human Skeletal Muscle

Published on: January 19, 2017

[Interference in CRP measurements in general practice].

Anne Lindegaard Christiansen1, Helena Strand Clemmensen1, Signe Bek Sørensen1

  • 1Blodprøver, Biokemi og Immunologi, Sygehus Sønderjylland, Syddansk Universitetshospital, Aabenraa.

Ugeskrift for Laeger
|July 3, 2026
PubMed
Summary
This summary is machine-generated.

Analytical interference can cause incorrect patient treatment. This case highlights how a plasma M-component interfered with a C-reactive protein (CRP) test on a point-of-care device, emphasizing the need for careful interpretation of results.

More Related Videos

Povidone Iodine Rectal Preparation at Time of Prostate Needle Biopsy is a Simple and Reproducible Means to Reduce Risk of Procedural Infection
05:32

Povidone Iodine Rectal Preparation at Time of Prostate Needle Biopsy is a Simple and Reproducible Means to Reduce Risk of Procedural Infection

Published on: September 21, 2015

Related Experiment Videos

Last Updated: Jul 4, 2026

Phosphorus-31 Magnetic Resonance Spectroscopy: A Tool for Measuring In Vivo Mitochondrial Oxidative Phosphorylation Capacity in Human Skeletal Muscle
09:40

Phosphorus-31 Magnetic Resonance Spectroscopy: A Tool for Measuring In Vivo Mitochondrial Oxidative Phosphorylation Capacity in Human Skeletal Muscle

Published on: January 19, 2017

Povidone Iodine Rectal Preparation at Time of Prostate Needle Biopsy is a Simple and Reproducible Means to Reduce Risk of Procedural Infection
05:32

Povidone Iodine Rectal Preparation at Time of Prostate Needle Biopsy is a Simple and Reproducible Means to Reduce Risk of Procedural Infection

Published on: September 21, 2015

Area of Science:

  • Clinical Biochemistry
  • Analytical Chemistry
  • Point-of-Care Testing

Background:

  • Analytical interference is a significant challenge in clinical biochemistry.
  • Unidentified interference can lead to inaccurate patient results and inappropriate medical treatment.
  • Point-of-care devices are increasingly used in general practice, making interference detection crucial.

Purpose of the Study:

  • To report a case of analytical interference affecting a C-reactive protein (CRP) result.
  • To illustrate the potential impact of plasma M-components on point-of-care testing.
  • To emphasize the importance of considering analytical interference in clinical practice.

Main Methods:

  • A case study approach was used.
  • A C-reactive protein (CRP) analysis was performed using a point-of-care device.
  • Investigation into the cause of interference involved detecting a plasma M-component.

Main Results:

  • Analytical interference was observed for a CRP result.
  • The interference was attributed to the presence of a plasma M-component.
  • The interference led to a falsely reported CRP value.

Conclusions:

  • Plasma M-components can cause analytical interference in point-of-care CRP testing.
  • Awareness of potential interferences is vital for accurate interpretation of biochemical results.
  • Clinicians should consider analytical interference when results appear inconsistent with the clinical picture.