Closed-loop correction of illumination-focal-plane mismatch enables consistent multicolor sectioning in OS-SIM
Related Experiment Video
Updated: Nov 1, 2026

Simultaneous Multicolor Imaging of Biological Structures with Fluorescence Photoactivation Localization Microscopy
Published on: December 9, 2013
Yuqi Qin1, Xinshuang Cao1, Ziyu Lei1
1MOE Key Laboratory for Biomedical Photonics, Wuhan National Laboratory for Optoelectronics, Huazhong University of Science and Technology, Wuhan 430074, China.
Abstract:
In multicolor optical-sectioning structured illumination microscopy (OS-SIM), wavelength-dependent illumination-focal-plane mismatch shifts the effective sectioning planes of different channels relative to a common detection focal plane, producing inconsistent sectioning performance. We present a closed-loop strategy that (i) localizes each channel's illumination-focal-plane position from the central peak of its axial modulation response, (ii) aligns the planes with channel-specific equal-focal-length 4f modules, and (iii) verifies residual mismatch by PZT-driven objective scanning. In our three-color system, the blue illumination focal plane was displaced 13--14 µm from both the red and green planes before correction; after correction, the median absolute peak-position differences |R-G|, |R-B|, and |G-B| fell to 0.25, 0.20, and 0.11 µm, respectively. Fluorescent-bead and brain-slice imaging confirmed improved interchannel sectioning consistency and fewer associated artifacts. The approach directly corrects illumination-focal-plane mismatch, a physical source of interchannel sectioning inconsistency, rather than relying solely on detection-side chromatic aberration correction or postprocessing registration.
Related Concept Videos
Confocal Fluorescence Microscopy
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...