Related Experiment Videos

Effect of a Millipore filter on complications of intravenous infusions: a prospective clinical trial

British Medical Journal
|November 24, 1973
PubMed

Insights

The Millipore filter unit did not reduce bacterial contamination or thrombophlebitis during intravenous infusions. This filter also caused flow issues and blockages, limiting its clinical use.

Area of Science:

  • Medical Devices
  • Infection Control
  • Intravenous Therapy

Background:

  • Intravenous (IV) infusions can introduce bacteria into circulation.
  • Millipore filter units were proposed to mitigate this risk.
  • Thrombophlebitis is a common complication of IV therapy.

Purpose of the Study:

  • To evaluate the efficacy of Millipore filter units in reducing bacterial contamination during IV infusions.
  • To assess the impact of Millipore filters on the incidence of thrombophlebitis.
  • To determine the practical usability and performance of the filter units.

Main Methods:

  • A prospective study was conducted to assess filter performance.
  • In-vitro and in-vivo experiments were performed.
  • Bacterial contamination of cannulae and thrombophlebitis incidence were monitored.

Main Results:

  • No significant reduction in thrombophlebitis or cannula bacterial contamination was observed.
  • The filter units demonstrated inconvenient handling and reduced flow rates.
  • Frequent filter blockages and incompatibility with blood/fat emulsions were noted.

Conclusions:

  • Millipore filter units are ineffective in preventing bacterial contamination during IV infusions.
  • The filters present significant practical limitations and performance issues.
  • Their use in routine intravenous therapy is not recommended due to these drawbacks.