Corrosion of pacemaker electrodes.

V Parsonnet, A Villanueva, J Driller

    Pacing and Clinical Electrophysiology : PACE
    |May 1, 1981
    PubMed
    Summary
    This summary is machine-generated.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Effects of time-restricted eating and resistance training on skeletal muscle tissue quantity, quality and function in postmenopausal women with overweight or obesity: A study protocol.

    Nutrition, metabolism, and cardiovascular diseases : NMCD·2025
    Same author

    Bispectral index monitoring for sedation in scheduled adult colonoscopy: A randomized controlled trial.

    Revista espanola de anestesiologia y reanimacion·2024
    Same author

    Fluorescence labeling of mitochondria in living cells by the cationic photosensitizer ZnTM2,3PyPz, and the possible roles of redox processes and pseudobase formation in facilitating dye uptake.

    Biotechnic & histochemistry : official publication of the Biological Stain Commission·2022
    Same author

    Risk factors for clinical deterioration in patients admitted for COVID-19: A case-control study.

    Revista clinica espanola·2021
    Same author

    [Risk factors for clinical deterioration in patients admitted for COVID-19: A case-control study].

    Revista clinica espanola·2021
    Same author

    Evaluation of oncogenic cysteinyl leukotriene receptor 2 as a therapeutic target for uveal melanoma.

    Cancer metastasis reviews·2018
    Same journal

    A Case of Permanent Pacemaker Implantation via the Epicardial Approach Using the 3830 Lead in an 11-Day-Old Neonate (With Follow-Up of the Above Case).

    Pacing and clinical electrophysiology : PACE·2026
    Same journal

    Cryoballoon Versus Radiofrequency Ablation for Persistent Atrial Fibrillation: Meta-Analysis of Randomized Trials.

    Pacing and clinical electrophysiology : PACE·2026
    Same journal

    Tilt Test Duration in Suspected Vasovagal Syncope: Temporal Patterns and Diagnostic Yield in Patients From Central China.

    Pacing and clinical electrophysiology : PACE·2026
    Same journal

    Combined Leadless Pacing and Subcutaneous ICD Therapy in Long QT Syndromes.

    Pacing and clinical electrophysiology : PACE·2026
    Same journal

    Association of Anesthesia Modality With Procedural Parameters and Clinical Outcomes in PVI for Atrial Fibrillation.

    Pacing and clinical electrophysiology : PACE·2026
    Same journal

    Zero-Fluoroscopy Cryoballoon Ablation for Paroxysmal Atrial Fibrillation in a Patient With Dextrocardia: A Case Report.

    Pacing and clinical electrophysiology : PACE·2026
    See all related articles

    Pacemaker electrode corrosion is microscopic and linked to implant duration and pacemaker DC offset currents. Most corroded electrodes did not cause adverse clinical effects.

    Area of Science:

    • Biomaterials Science
    • Medical Device Engineering
    • Cardiology

    Background:

    • Pacemaker longevity has increased, shifting focus to component durability.
    • Electrode corrosion is an emerging concern in long-term pacemaker function.

    Purpose of the Study:

    • To investigate the prevalence and characteristics of pacemaker electrode corrosion.
    • To correlate corrosion with explantation reasons, pacemaker status, and implant duration.

    Main Methods:

    • Examination of 56 explanted pacemaker electrodes using scanning electron microscopy.
    • Grading corrosion severity using a numerical system.
    • Correlation with explantation data, pacemaker condition, and DC offset currents.

    Main Results:

    Related Experiment Videos

    • Microscopic corrosion was observed on nearly all explanted electrodes.
    • Significant corrosion correlated with DC offset currents > 5 microamperes.
    • Corrosion severity increased with longer implantation durations.

    Conclusions:

    • Pacemaker electrode corrosion is common, microscopic, and related to implant duration and electrical factors.
    • Clinical adverse effects were minimal even with significant corrosion.
    • Long-term clinical impact of electrode corrosion remains uncertain.