Coronary sinus pacing clinical follow-up.
Circulation
|January 1, 1978
Summary
Coronary sinus pacing offers a safe and effective method for atrial pacing. This study found a 14% failure rate and 14% transient issues, with no pacemaker-related deaths in 66 patients.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Devices
Background:
- Atrial pacing is crucial for managing bradyarrhythmias.
- Coronary sinus (CS) pacing presents an alternative to traditional atrial pacing.
- Evaluating the safety and efficacy of CS pacing is essential.
Purpose of the Study:
- To assess the safety and effectiveness of coronary sinus pacing.
- To determine the feasibility of achieving adequate atrial pacing thresholds.
- To evaluate CS pacing in patients with atrioventricular block.
Main Methods:
- Retrospective analysis of 66 patients undergoing coronary sinus pacing.
- Follow-up averaged 14 months to assess pacing parameters and complications.
- Evaluation of pacing thresholds and success rates in patients with AV block.
Main Results:
- Coronary sinus pacing demonstrated a 14% failure rate and 14% transient complications.
- Successful atrial pacing was achieved with thresholds up to 6.0 mA.
- 18 of 19 patients with atrioventricular block showed successful atrial pacing.
Conclusions:
- Coronary sinus pacing is a safe and effective strategy for atrial pacing.
- The technique is feasible even in patients with challenging atrioventricular conduction.