Optimized assay for serum angiotensin-converting enzyme activity.
Clinical Chemistry
|December 1, 1981
Summary
This study presents an optimized assay for measuring serum angiotensin-converting enzyme (ACE) activity. The new colorimetric method offers reliable results, unaffected by common interferences, establishing new reference intervals for men and women.
Area of Science:
- Clinical Chemistry
- Enzyme Assays
- Biochemical Diagnostics
Background:
- Serum angiotensin-converting enzyme (ACE) levels are crucial biomarkers in various physiological and pathological conditions.
- Existing assays may be subject to interferences or lack optimal precision.
- A need exists for a robust and reliable method for quantifying serum ACE activity.
Purpose of the Study:
- To develop and validate an optimized two-point kinetic assay for serum angiotensin-converting enzyme (ACE).
- To establish the performance characteristics and reference intervals for the new assay.
Main Methods:
- The assay utilizes a colorimetric determination of hippurate released from hippuryl-L-histidyl-L-leucine.
- Angiotensin-converting enzyme (ACE) activity is measured in the presence of chloride ions.
- Assay validation included assessment of precision, linearity, and interference from common substances like lipemia, icterus, and hemoglobin.
Main Results:
- The optimized assay demonstrated excellent precision, with within-run and between-run coefficients of variation (CVs) ranging from 2.1% to 3.2%.
- Linearity was confirmed up to 200 U/L, indicating a wide dynamic range.
- The method showed minimal interference from lipemia and icterus, though high hemoglobin concentrations (>1.5 g/L) caused slight negative interference.
- Reference intervals were established as 22-82 U/L for men and 25-69 U/L for women.
Conclusions:
- The developed two-point kinetic assay provides a precise and reliable method for quantifying serum angiotensin-converting enzyme (ACE) activity.
- The assay's robustness against common interfering substances enhances its clinical utility.
- Established reference intervals aid in the accurate interpretation of serum ACE levels in patient populations.