Oxygen tension measurements: a comparison of methodologies
American Journal of Optometry and Physiological Optics
|August 1, 1984
Summary
The equivalent oxygen percentage (EOP) technique and Rasson and Fatt method provide similar oxygen values under contact lenses for 0.23 mm thick lenses. Thicker or thinner lenses show discrepancies, but a linear relationship exists between center thickness and the ratio of oxygen values.
Area of Science:
- Ophthalmology
- Biomedical Engineering
- Optometry
Background:
- Quantifying oxygen tension under contact lenses is crucial for eye health.
- The equivalent oxygen percentage (EOP) technique and Rasson and Fatt method are common approaches for open-eyelid conditions.
- Variations in contact lens properties may affect oxygen tension measurements.
Purpose of the Study:
- To compare the EOP technique and the Rasson and Fatt method for measuring sub-contact lens oxygen tension.
- To investigate the influence of contact lens center thickness and oxygen transmissibility on these measurements.
Main Methods:
- Five subjects wore six different contact lenses with varying center thicknesses and oxygen transmissibilities.
- Oxygen tension was measured using both the EOP technique and the Rasson and Fatt method for each lens.
- Statistical analysis was performed to compare the results and identify relationships.
Main Results:
- Both methods yielded similar oxygen values for contact lenses with a center thickness of approximately 0.23 mm.
- Significant differences in oxygen values were observed for lenses considerably thicker or thinner than 0.23 mm.
- A strong linear correlation was identified between contact lens center thickness and the ratio of oxygen values obtained by the two methods.
Conclusions:
- The choice of measurement technique for sub-contact lens oxygen tension may be critical, especially for lenses deviating from 0.23 mm center thickness.
- The observed linear relationship provides a potential method for reconciling discrepancies between the two techniques based on lens thickness.