A single indicator technique to estimate extravascular lung water.
The Journal of Surgical Research
|November 1, 1982
Summary
A new method using a single thermal indicator (SI) accurately estimates extravascular thermal volume (EVTV). This technique aligns with gravimetric and double indicator (DI) dilution methods for measuring extravascular lung water (EVLW).
Area of Science:
- Physiology
- Medical Measurement
- Cardiovascular Research
Background:
- Accurate measurement of extravascular thermal volume (EVTV) is crucial for clinical assessment.
- Existing methods like double indicator (DI) dilution or gravimetric analysis can be invasive or complex.
Purpose of the Study:
- To evaluate the accuracy of a single thermal indicator (SI) method for estimating extravascular thermal volume (EVTV).
- To compare SI-derived EVTV with established methods like gravimetric measurements of extravascular lung water (EVLW) and DI dilution.
Main Methods:
- Utilized a single thermal indicator (SI) method with thermal dilution, sampling from pulmonary and femoral arteries.
- Compared SI-derived EVTV with direct gravimetric measurements of EVLW in 47 dogs.
- Validated SI-derived EVTV against simultaneous double indicator (DI) dilution measurements in 62 dogs and 38 humans.
Main Results:
- The SI method showed strong correlation with gravimetric EVLW measurements (r=0.91).
- SI-derived EVTV closely agreed with DI dilution measurements in both dogs (r=0.94) and humans (r=0.97).
- Measurements were performed without blood withdrawal using standard catheters.
Conclusions:
- Single thermal indicator (SI) measurements provide a reliable estimation of extravascular thermal volume (EVTV).
- This method offers a less invasive and practical alternative to existing techniques for assessing extravascular lung water (EVLW).