A simple, inexpensive quality control material for ortho ELT-8 platelet counts
Annals of Clinical Biochemistry
|March 1, 1983
Summary
Polystyrene latex microspheres offer a simple and affordable method for quality control in platelet counting. These materials provide reliable results comparable to commercial controls for the Ortho ELT-8 analyzer.
Area of Science:
- Hematology
- Clinical Chemistry
- Materials Science
Background:
- Accurate platelet counting is crucial for diagnosing and managing various medical conditions.
- Existing quality control methods can be expensive or complex.
- There is a need for simple, cost-effective quality control materials for platelet counters.
Purpose of the Study:
- To evaluate the efficacy of polystyrene latex microsphere suspensions as quality control materials for the Ortho ELT-8 platelet counter.
- To assess the cost-effectiveness and simplicity of using these microspheres.
Main Methods:
- Utilizing 1 µm diameter polystyrene latex microsphere suspensions at dilutions of 1:600 and 1:3000.
- Performing quality control assessments over a 12-week period.
- Measuring the coefficient of variation (CV) at different platelet count levels (350 x 10^9/l and 70 x 10^9/l).
Main Results:
- The coefficient of variation (CV) at the 350 x 10^9/l level was 2.2% over 12 weeks.
- The CV at the 70 x 10^9/l level was 3.5%.
- Performance was comparable to a commercial quality control material.
Conclusions:
- Polystyrene latex microsphere suspensions are simple, inexpensive, and effective materials for the quality control of Ortho ELT-8 platelet counts.
- These microspheres provide reliable and reproducible results, offering a viable alternative to commercial controls.
- The study demonstrates the potential of these materials for routine laboratory use in hematology quality assurance.