Related Experiment Videos
Experience with semirigid rod and inflatable penile prostheses
The Journal of Urology
|May 1, 1983
Summary
Penile prosthesis surgery outcomes show semirigid rods had minimal complications, while inflatable prostheses required revisions, especially with lower abdominal incisions. Scrotal incisions improved outcomes for inflatable penile prostheses.
Area of Science:
- Urology
- Surgical Technology
- Medical Device Evaluation
Background:
- Penile prostheses are used to treat erectile dysfunction.
- Different types of penile prostheses exist, including semirigid rods and inflatable devices.
- Surgical approach and device design can influence complication rates.
Purpose of the Study:
- To evaluate the complication rates and revision requirements for different types of penile prostheses implanted between 1975 and 1980.
- To compare outcomes between semirigid rod prostheses and inflatable penile prostheses.
- To assess the impact of surgical incision site on inflatable penile prosthesis outcomes.
Main Methods:
- Retrospective review of 169 patients who received penile prostheses (Small-Carrion, Finney, Jonas, Scott) from 1975-1980.
- Categorization of prostheses into semirigid rod and inflatable types.
- Analysis of complication data, specifically urethral erosion and surgical revisions, based on prosthesis type and surgical approach (lower abdominal vs. scrotal incision).
Main Results:
- 48 patients received semirigid rod prostheses, with urethral erosion as the sole complication.
- 121 patients received inflatable prostheses.
- Of the inflatable prostheses, the first 70 implanted via a lower abdominal incision required 63 revisions in 34 patients; the subsequent 51 implanted via a scrotal incision with rear tip extenders required only 4 revisions.
Conclusions:
- Semirigid rod penile prostheses are associated with a low complication rate.
- Inflatable penile prostheses implanted via a lower abdominal incision have a high revision rate.
- Scrotal incision and the use of rear tip extenders appear to reduce the revision rate for inflatable penile prostheses.