Evaluation of three spirometers on healthy children
Acta Paediatrica Scandinavica
|November 1, 1978
Summary
Comparing spirometers in children, this study found electronic and bellows devices closely matched a standard for vital capacity (VC) and forced expiratory volume in 1 second (FEV1.0). However, peak expiratory flow rate (PEFR) measurements varied significantly between devices.
Area of Science:
- Pediatric Pulmonology
- Medical Device Evaluation
- Respiratory Physiology
Background:
- Spirometry is crucial for assessing lung function in children.
- Accurate spirometer calibration and comparison are essential for reliable clinical data.
- Various spirometer technologies exist, necessitating performance validation.
Purpose of the Study:
- To compare the accuracy of two electronic spirometers (Dräger Spirotron, Monaghan M403) and a wedge bellows spirometer (Vitalograph) against a Bernstein spirometer in healthy children.
- To evaluate the agreement of vital capacity (VC) and forced expiratory volume in 1 second (FEV1.0) measurements.
- To assess the reliability of peak expiratory flow rate (PEFR) measurements from electronic spirometers compared to a Wright peak flow meter.
Main Methods:
- A cohort of 61 healthy children (30 girls, 31 boys) underwent spirometry testing.
- Measurements of VC and FEV1.0 were compared using regression analysis against body height cubed.
- Paired difference analysis was performed to quantify measurement discrepancies.
- PEFR was measured using electronic devices and a Wright peak flow meter for comparison.
Main Results:
- Regression lines for VC and FEV1.0 showed high correlation coefficients and similar standard deviation values across all tested spirometers relative to body height.
- Slight differences in mean values were observed through paired difference analysis, with standard deviations for VC ranging from 4.6% to 6.6% and for FEV1.0 from 4.8% to 6.2%.
- PEFR values from the Dräger Spirotron and Monaghan M403 significantly deviated from those obtained by the Wright peak flow meter.
Conclusions:
- Electronic and wedge bellows spirometers demonstrate good agreement with a standard for measuring VC and FEV1.0 in children.
- The accuracy of PEFR measurement varies significantly between different electronic spirometer models and a standard peak flow meter.
- Careful selection of spirometry equipment is necessary, particularly for PEFR assessment, to ensure reliable pediatric respiratory function data.