High frequency Bekesy audiometry. I: equipment and calibration
Summary
This study details an affordable modification to the Grason-Stadler E-800 Bekesy audiometer, enabling high-frequency (HF) hearing tests up to 20 kHz. The enhanced audiometer offers valuable diagnostic insights for audiologists.
Area of Science:
- Audiology
- Medical instrumentation
- Hearing diagnostics
Background:
- The Grason-Stadler E-800 Bekesy audiometer is a standard tool for hearing assessments.
- Existing limitations may restrict its utility for high-frequency (HF) range testing.
- Cost-effective solutions are sought for expanding diagnostic capabilities in audiology.
Purpose of the Study:
- To describe a simple, inexpensive conversion of the Grason-Stadler E-800 Bekesy audiometer.
- To extend the high-frequency (HF) range of the audiometer up to 20 kHz.
- To provide a cost-effective method for obtaining HF diagnostic information.
Main Methods:
- Modification of the Grason-Stadler E-800 Bekesy audiometer.
- Incorporation of a 1/2-inch condenser microphone as a transducer.
- Utilizing an adjustable headband for coupling the microphone to the ear canal.
- Free field calibration recommended for accuracy.
- Detailed instructions for conversion and calibration provided.
Main Results:
- The modified system achieves a dynamic range of 135 dB.
- Prototype unit calibrations demonstrated acceptable stability and validity.
- The conversion is straightforward and inexpensive.
Conclusions:
- The modified HF E-800 audiometer is a valuable tool for audiologists.
- It provides a cost-effective source of high-frequency diagnostic information.
- This modification enhances the utility of existing audiometric equipment.