Quantifying female incontinence with particular reference to the Urilos System
Urologia Internationalis
|January 1, 1980
Summary
The Urilos System shows poor reliability for quantifying urinary incontinence due to significant variations in urine loss measurements between nappy packs. Alternative objective methods, like the daily bladder chart, are recommended for accurate assessment.
Area of Science:
- Urology
- Medical Devices
- Clinical Assessment
Background:
- Urinary incontinence quantification is crucial for patient management.
- Existing methods for measuring urine loss may have limitations.
- The Urilos System offers a potential new approach to quantify incontinence.
Purpose of the Study:
- To assess the reliability and accuracy of the Urilos System for quantifying urinary incontinence.
- To evaluate the reproducibility of urine loss measurements using the Urilos System.
- To compare the performance of the Urilos System across different nappy packs.
Main Methods:
- The Urilos System was used to measure urine loss.
- Reproducibility was assessed within the same pack of nappies.
- Comparisons were made between different packs of nappies.
Main Results:
- Approximate reproducibility of urine loss measurements was found within the same nappy pack.
- Under- and over-recordings of urine loss up to 25% and 23% were observed, respectively.
- Differences in measurements up to 68% were noted when comparing different nappy packs.
Conclusions:
- The Urilos System currently lacks sufficient reliability for accurate incontinence quantification.
- Significant variability exists in measurements between different nappy packs.
- The daily bladder chart is suggested as a more reliable method for objective incontinence measurement.