Transferability of blood lead determinations by furnace atomic absorption spectrophotometry and continuum background
H Qiao1, P J Parsons, W Slavin
1Wadsworth Center, New York State Department of Health, Albany, NY 12201, USA.
Abstract:
We have examined and proved feasible the transfer of a method for blood lead determination, developed and optimized for a Zeeman-corrected instrument, to a continuum-corrected furnace atomic absorption spectrophotometer. Numerous reference materials analyzed with the continuum-corrected instrument gave results within 10 micrograms/L (0.05 mumol/L) at low values and varied by < 6% at values > 200 micrograms/L (0.97 mumol/L). Forty-four routine human blood specimens were analyzed by the same method with both continuum- and Zeeman-corrected instrumentation, and gave results that agreed within about the same limits as found with reference materials. The day-to-day precision was about 1/5 the accuracy results. The detection limit was approximately 5 micrograms/L (0.025 mumol/L).
Related Concept Videos
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Lab