X-ray imaging using amorphous selenium: inherent spatial resolution

W Que1, J A Rowlands

  • 1Sunnybrook Health Science Centre, University of Toronto, Ontario, Canada.

Medical Physics
|April 1, 1995
PubMed
Summary

This study analyzes amorphous selenium (a-Se) plates for diagnostic X-ray imaging. Amorphous selenium plates offer superior inherent spatial resolution compared to cesium iodide layers, with photoelectron ranges and X-ray incidence being key limiting factors.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Particle size-dependent bioaccessibility of chromium in smelting soils: assessment by multi-method in vitro simulations.

Environmental monitoring and assessment·2025
Same author

Sci-Fri AM: Imaging - 02: High resolution detectors for PET mammography.

Medical physics·2017
Same author

Poster - Thurs Eve-16: Just-in-time tomography (JiTT).

Medical physics·2017
Same author

Sci-Fri AM: YIS-10: Development of a flat panel detector with avalanche gain for low-dose x-ray imaging.

Medical physics·2017
Same author

Magnetic diagnostics for equilibrium reconstruction and realtime plasma control in NSTX-Upgrade.

The Review of scientific instruments·2014
Same author

Monte Carlo simulation of a quantum noise limited Čerenkov detector based on air-spaced light guiding taper for megavoltage x-ray imaging.

Medical physics·2014

Area of Science:

  • Medical Physics
  • Radiological Imaging Technology
  • Materials Science

Background:

  • Diagnostic X-ray imaging relies on high spatial resolution for accurate image interpretation.
  • Amorphous selenium (a-Se) plates are utilized in direct radiography systems.
  • Understanding the factors limiting spatial resolution in a-Se is crucial for optimizing imaging performance.

Purpose of the Study:

  • To theoretically investigate the inherent spatial resolution of amorphous selenium plates.
  • To identify and quantify the physical effects that degrade spatial resolution in a-Se imaging.
  • To compare the resolution capabilities of a-Se with other detector materials like cesium iodide.

Main Methods:

  • Theoretical analysis of spatial resolution limitations.

Related Experiment Videos

  • Consideration of primary photoelectron ranges.
  • Evaluation of K fluorescence and Compton scattered photon reabsorption.
  • Assessment of diffusion, geometric, electrostatic, and space charge effects.
  • Estimation of the modulation transfer function (MTF) for a-Se.
  • Main Results:

    • Primary photoelectron ranges and geometric effects from oblique X-ray incidence were identified as the primary resolution-limiting factors.
    • The modulation transfer function of a-Se was estimated for the diagnostic X-ray energy range.
    • The theoretical inherent spatial resolution of a-Se plates was found to be significantly better than that of cesium iodide layers.

    Conclusions:

    • Amorphous selenium exhibits excellent inherent spatial resolution for diagnostic X-ray imaging.
    • The study provides a theoretical framework for understanding resolution limitations in a-Se detectors.
    • a-Se technology presents a promising alternative for high-resolution X-ray imaging applications.