Recovery characteristics of a rigid, nonmetallic microdialysis probe for use in an electromagnetic field
1Operational Technologies Corporation, San Antonio, Texas, USA.
Abstract:
It is well known that metal objects perturb electromagnetic fields. Therefore, a conventional metal microdialysis probe cannot be used to determine the bioeffects of electromagnetic radiation. Using fused-silica tubing, we developed an inexpensive nonmetallic, rigid microdialysis probe for use in electromagnetic radiation research or during magnetic resonance imaging. This probe has a concentric tube design, with the membrane length adjustable to the size of the area to be dialyzed. The probes tested had regenerated-cellulose membranes that were 3 mm in length. This report describes how to make this probe. Average relative recovery rates at flow rates of 2.0, 1.0, and 0.5 microliters/min were 21%, 27%, and 42%, respectively. These rates were slightly lower than the 30%, 42%, and 68% obtained with the commercially available metallic CMA10 microdialysis probe with a 3 mm membrane. This may be due to the fused-silica probe and CMA10 probe being made with different types of dialysis membranes.