On-line cardiac output by digital computer
Journal of Applied Physiology
|February 1, 1976
Summary
This study presents an on-line computer system for accurate cardiac output calculation using the Stewart-Hamilton method. The system minimizes errors from assumptions, improving precision in dye-dilution curve analysis.
Area of Science:
- Cardiovascular Physiology
- Medical Instrumentation
- Computational Biology
Background:
- Conventional cardiac output devices often deviate from the Stewart-Hamilton method, leading to significant errors.
- Accurate measurement of cardiac output is crucial for diagnosing and managing various cardiovascular conditions.
Purpose of the Study:
- To develop and validate an on-line computer system for precise cardiac output calculation using the classical Stewart-Hamilton method.
- To address and mitigate errors introduced by common assumptions in existing cardiac output measurement devices.
Main Methods:
- Implemented a digital filter to smooth dye-dilution curves and eliminate heartbeat noise.
- Utilized the exponential property of recirculation for accurate removal.
- Developed quality control metrics including baseline length, exponential fit, and a figure of merit based on the five tanks-in-series model.
Main Results:
- The on-line system demonstrated high accuracy, with a mean difference of only 150 ml compared to hand calculations.
- For simulated dye curves, the computer system achieved a standard deviation of 10 ml, significantly outperforming hand calculations (SD = 200 ml).
- Visual checks via CRT display of curves and parameters ensured data integrity.
Conclusions:
- The developed on-line system provides a reliable and accurate method for cardiac output determination via dye-dilution.
- This system offers a significant improvement in precision and error reduction compared to traditional methods and manual calculations.