[False sensitivity of Enterococcus faecium to amoxicillin-clavulanic acid with the MicroScan system]

P Joyanes1, A Pascual, A I Suárez

  • 1Departamento de Microbiología, Hospital Universitario Virgen Macarena, Sevilla.

Abstract

Insights

MicroScan system shows discrepancies in Enterococcus faecium susceptibility testing. For ampicillin-resistant strains, consider them resistant to amoxicillin-clavulanic acid, disregarding MicroScan results.

Area of Science:

  • Clinical microbiology
  • Antimicrobial susceptibility testing

Background:

  • Discrepancies observed with the MicroScan system for Enterococcus faecium susceptibility to ampicillin and amoxicillin/clavulanic acid.
  • Enterococcus faecium is an important human pathogen.

Purpose of the Study:

  • To evaluate discrepancies in amoxicillin/clavulanic acid susceptibility testing of Enterococcus faecium using the MicroScan system compared to standard methods.
  • To determine the reliability of MicroScan for predicting amoxicillin/clavulanic acid resistance in E. faecium.

Main Methods:

  • Susceptibility testing of 76 Enterococcus faecium strains using the MicroScan automated system.
  • Comparison of MicroScan results with standard microdilution assays for ampicillin, amoxicillin, and amoxicillin/clavulanic acid.
  • Evaluation of the impact of bacterial inoculum and betalactamase production on MIC values.

Main Results:

  • MicroScan identified 98.7% of E. faecium strains as inhibited by amoxicillin-clavulanic acid (16 mg/l) versus 60.5% for ampicillin.
  • MIC values for amoxicillin/clavulanic acid were significantly lower with MicroScan (< or = 4/2 mg/l) compared to standard microdilution (> or = 16/8 mg/l) for most strains.
  • The observed discrepancies were independent of betalactamase production or bacterial inoculum size.

Conclusions:

  • The MicroScan system may underestimate amoxicillin-clavulanic acid resistance in Enterococcus faecium.
  • Enterococcus faecium strains found resistant to ampicillin (non-betalactamase producers) should be considered resistant to amoxicillin-clavulanic acid, irrespective of MicroScan-generated values.
  • Clinical laboratories should exercise caution when interpreting MicroScan results for amoxicillin-clavulanic acid susceptibility in E. faecium.